Detailed Information

Cited 2 time in webofscience Cited 3 time in scopus
Metadata Downloads

Sliding Time of Flight: Sliding Time of Flight MR Angiography Using a Dynamic Image Reconstruction Method

Full metadata record
DC Field Value Language
dc.contributor.authorChoi, Joonsung-
dc.contributor.authorSeo, Hyunseok-
dc.contributor.authorLim, Yongwan-
dc.contributor.authorHan, Yeji-
dc.contributor.authorPark, HyunWook-
dc.date.available2020-02-28T09:47:04Z-
dc.date.created2020-02-06-
dc.date.issued2015-03-
dc.identifier.issn0740-3194-
dc.identifier.urihttps://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/10719-
dc.description.abstractPurposeTo obtain three-dimensional (3D) MR angiography having high contrast between vessel and stationary background tissue, a novel technique called sliding time of flight (TOF) is proposed. MethodsThe proposed method relies on the property that flow-related enhancement (FRE) is maximized at the blood-entering slice in an imaging slab. For the proposed sliding TOF, a sliding stack-of-stars sampling and a dynamic MR image reconstruction algorithm were developed. To verify the performance of the proposed method, in vivo study was performed and the results were compared with multiple overlapping thin 3D slab acquisition (MOTSA) and sliding interleaved k(y) (SLINKY). ResultsIn MOTSA and SLINKY, the variation of FRE resulted in severe venetian blind (MOTSA) or ghost (SLINKY) artifacts, while the vessel-contrast increased as the flip angle of radiofrequency (RF) pulses increased. On the other hand, the proposed method could provide high-contrast angiograms with reduced FRE-related artifacts. ConclusionThe sliding TOF can provide 3D angiography without image artifacts even if high flip angle RF pulses with thick slab excitation are used. Although remains of subsampling artifacts can be present in the reconstructed images, they can be reduced by MIP operation and resolved further by regularization techniques. Magn Reson Med 73:1177-1183, 2015. (c) 2014 Wiley Periodicals, Inc.-
dc.language영어-
dc.language.isoen-
dc.publisherWILEY-BLACKWELL-
dc.relation.isPartOfMAGNETIC RESONANCE IN MEDICINE-
dc.subjectTHIN SLAB ACQUISITION-
dc.subjectARBITRARY K-SPACE-
dc.subjectMAGNETIZATION-TRANSFER-
dc.titleSliding Time of Flight: Sliding Time of Flight MR Angiography Using a Dynamic Image Reconstruction Method-
dc.typeArticle-
dc.type.rimsART-
dc.description.journalClass1-
dc.identifier.wosid000350279900031-
dc.identifier.doi10.1002/mrm.25215-
dc.identifier.bibliographicCitationMAGNETIC RESONANCE IN MEDICINE, v.73, no.3, pp.1177 - 1183-
dc.identifier.scopusid2-s2.0-84923209502-
dc.citation.endPage1183-
dc.citation.startPage1177-
dc.citation.titleMAGNETIC RESONANCE IN MEDICINE-
dc.citation.volume73-
dc.citation.number3-
dc.contributor.affiliatedAuthorHan, Yeji-
dc.type.docTypeArticle-
dc.subject.keywordAuthorMOTSA-
dc.subject.keywordAuthorstack-of-stars trajectory-
dc.subject.keywordAuthorsliding slab-
dc.subject.keywordAuthorTOF-
dc.subject.keywordPlusTHIN SLAB ACQUISITION-
dc.subject.keywordPlusARBITRARY K-SPACE-
dc.subject.keywordPlusMAGNETIZATION-TRANSFER-
dc.relation.journalResearchAreaRadiology, Nuclear Medicine & Medical Imaging-
dc.relation.journalWebOfScienceCategoryRadiology, Nuclear Medicine & Medical Imaging-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
Files in This Item
There are no files associated with this item.
Appears in
Collections
보건과학대학 > 의용생체공학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Han, Ye Ji photo

Han, Ye Ji
College of IT Convergence (의공학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE