Detailed Information

Cited 3 time in webofscience Cited 4 time in scopus
Metadata Downloads

Evaluation of automatic exposure control system chamber for the dose optimization when examining pelvic in digital radiography

Authors
Kim, Sung-ChulLee, Hae-KagLee, Yang-SubCho, Jae-Hwan
Issue Date
Jun-2015
Publisher
IOS PRESS
Keywords
Digital radiography; auto exposure control; chamber; dose optimization
Citation
JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, v.23, no.3, pp.321 - 330
Journal Title
JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY
Volume
23
Number
3
Start Page
321
End Page
330
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/11925
DOI
10.3233/XST-150492
ISSN
0895-3996
Abstract
We found a way to optimize the image quality and reduce the exposure dose of patients through the proper activity combination of the automatic exposure control system chamber for the dose optimization when examining the pelvic anteroposterior side using the phantom of the human body standard model. We set 7 combinations of the chamber of automatic exposure control system. The effective dose was yielded by measuring five times for each according to the activity combination of the chamber for the dose measurement. Five radiologists with more than five years of experience evaluated the image through picture archiving and communication system using double blind test while classifying the 6 anatomical sites into 3-point level (improper, proper, perfect). When only one central chamber was activated, the effective dose was found to be the highest level, 0.287 mSv; and lowest when only the top left chamber was used, 0.165 mSv. After the subjective evaluation by five panel members on the pelvic image was completed, there was no statistically meaningful difference between the 7 chamber combinations, and all had good image quality. When testing the pelvic anteroposterior side with digital radiography, we were able to reduce the exposure dose of patients using the combination of the top right side of or the top two of the chamber.
Files in This Item
There are no files associated with this item.
Appears in
Collections
보건과학대학 > 방사선학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Sung Chul photo

Kim, Sung Chul
Health Science (Dept.of Radiology)
Read more

Altmetrics

Total Views & Downloads

BROWSE