Development of a Beam Stop Array System with Dual Scan Mode for Scatter Correction of Cone-beam CT

Authors
Ahn, So HyunChoi, JinhoLee, Kyu ChanKim, SiyongLee, Rena
Issue Date
Apr-2014
Publisher
KOREAN PHYSICAL SOC
Keywords
Cone-beam CT; Scattered X-ray; Image quality; Beam stop array; Monte Carlo simulation
Citation
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, v.64, no.8, pp.1220 - 1229
Journal Title
JOURNAL OF THE KOREAN PHYSICAL SOCIETY
Volume
64
Number
8
Start Page
1220
End Page
1229
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/12759
DOI
10.3938/jkps.64.1220
ISSN
0374-4884
Abstract
A beam stop array (BSA) is widely used to estimate the scatter distribution of cone-beam computed tomography (CT) experimentally. A rectangular BSA, which is the most generally used, produces a penumbra region at the edge of the grid on the projection image. The penumbra region can be the cause of an inaccurate scatter estimate and loss of image data. In this study, two types of BSAs have been designed and developed, a linear BSA (l-BSA), which is a simple rectangular type, and a curved BSA (c-BSA) to avoid the penumbra effect. A Monte Carlo simulation was performed to determine the thickness and the material of beam stop arrays. The particle tracking technique of MCNP5 code was used to analyze the accuracy of the scatter distribution acquired by the beam stop arrays. The improvement in the image quality was analyzed by using an in-house phantom, a simple type CT phantom, and a humanoid phantom. How the curved beam stop array improved the penumbra effect was also analyzed in a large field of view. The additional dose caused by a dual scan was analyzed by using a Monte Carlo simulation and experiment. The accuracies of the scatter estimates were 83.4% for the c-BSA and 90.4% for the l-BSA. The c-BSA compensated the penumbra effect effectively while the l-BSA had a severe penumbra region in the partial projection image in a large field of view. Contrasts were improved by 24.8 (air-PMMA), 55.8 (Teflon-PMMA), and 81.7% (water-PMMA) for the c-BSA and 11.1 (air-PMMA), 44.1 (Teflon-PMMA), and 82.2% (water-PMMA) for the l-BSA by scatter correction. After noise suppression, the contrast-to-noise ratios were improved by 2.7-4.1 times for the c-BSA and 3.5-5.3 times for the l-BSA. Uniformities were also improved by up to 10 times for the c-BSA and 12 times for the l-BSA. The quality of the humanoid phantom image was also improved after the scatter correction. Consequently, the c-BSA can improve the image quality of cone-beam CT and compensate for the penumbra effect without an additional dose of radiation.
Files in This Item
There are no files associated with this item.
Appears in
Collections
의과대학 > 의학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Kyu Chan photo

Lee, Kyu Chan
College of Medicine (Department of Medicine)
Read more

Altmetrics

Total Views & Downloads

BROWSE