Detailed Information

Cited 79 time in webofscience Cited 87 time in scopus
Metadata Downloads

Vertical flow immunoassay (VFA) biosensor for a rapid one-step immunoassay

Full metadata record
DC Field Value Language
dc.contributor.authorOh, Young Kyoung-
dc.contributor.authorJoung, Hyou-Arm-
dc.contributor.authorKim, Sanghyo-
dc.contributor.authorKim, Min-Gon-
dc.date.available2020-02-29T04:42:17Z-
dc.date.created2020-02-06-
dc.date.issued2013-
dc.identifier.issn1473-0197-
dc.identifier.urihttps://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/15861-
dc.description.abstractA highly rapid, one-step immunoassay of high sensitivity C-reactive protein (hsCRP) using a biosensor with a vertical flow immunoassay (VFA) was developed. The VFA biosensor was primarily composed of a sample pad, conjugate pad, FTH film and nitrocellulose (NC) membrane, which were all vertically stacked upon one another. Anti-hsCRP and secondary antibodies were consecutively immobilized on the NC membrane at the position below the holes. Gold nanoparticles (AuNPs) conjugated with another anti-hsCRP antibody were encapsulated in the conjugation pad. Various assay conditions, including the size of the hole and the sample volume, were optimized. Under optimized conditions, hsCRP concentrations from 0.01 to 10 mg mL(-1) were detected within 2 min. In comparison with a lateral flow assay (LFA) system, the VFA sensor showed a gradual increase of signal in a concentration-dependent manner without a hook effect in the tested range.-
dc.language영어-
dc.language.isoen-
dc.publisherROYAL SOC CHEMISTRY-
dc.relation.isPartOfLAB ON A CHIP-
dc.subjectPOINT-OF-CARE-
dc.subjectC-REACTIVE PROTEIN-
dc.subjectTROPONIN-I-
dc.subjectBIOMARKERS-
dc.subjectDISEASE-
dc.subjectASSAY-
dc.titleVertical flow immunoassay (VFA) biosensor for a rapid one-step immunoassay-
dc.typeArticle-
dc.type.rimsART-
dc.description.journalClass1-
dc.identifier.wosid000314651000002-
dc.identifier.doi10.1039/c2lc41016h-
dc.identifier.bibliographicCitationLAB ON A CHIP, v.13, no.5, pp.768 - 772-
dc.identifier.scopusid2-s2.0-84875867885-
dc.citation.endPage772-
dc.citation.startPage768-
dc.citation.titleLAB ON A CHIP-
dc.citation.volume13-
dc.citation.number5-
dc.contributor.affiliatedAuthorKim, Sanghyo-
dc.type.docTypeArticle-
dc.subject.keywordPlusPOINT-OF-CARE-
dc.subject.keywordPlusC-REACTIVE PROTEIN-
dc.subject.keywordPlusTROPONIN-I-
dc.subject.keywordPlusBIOMARKERS-
dc.subject.keywordPlusDISEASE-
dc.subject.keywordPlusASSAY-
dc.relation.journalResearchAreaBiochemistry & Molecular Biology-
dc.relation.journalResearchAreaChemistry-
dc.relation.journalResearchAreaScience & Technology - Other Topics-
dc.relation.journalWebOfScienceCategoryBiochemical Research Methods-
dc.relation.journalWebOfScienceCategoryChemistry, Multidisciplinary-
dc.relation.journalWebOfScienceCategoryChemistry, Analytical-
dc.relation.journalWebOfScienceCategoryNanoscience & Nanotechnology-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
Files in This Item
There are no files associated with this item.
Appears in
Collections
바이오나노대학 > 바이오나노학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Sang Hyo photo

Kim, Sang Hyo
BioNano Technology (Department of BioNano Technology)
Read more

Altmetrics

Total Views & Downloads

BROWSE