A design of brain port with a built-in bi-directional check valve improves the ventriculostomy
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Chung, Taishin | - |
dc.contributor.author | Yoon, Ki-Cheol | - |
dc.contributor.author | Kim, Kwang Gi | - |
dc.contributor.author | Lee, Seung Hoon | - |
dc.contributor.author | Yoo, Heon | - |
dc.date.available | 2020-03-03T06:48:08Z | - |
dc.date.created | 2020-02-24 | - |
dc.date.issued | 2020-05 | - |
dc.identifier.issn | 0391-3988 | - |
dc.identifier.uri | https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/17849 | - |
dc.description.abstract | The increase of intracranial pressure is a life-threatening condition which requires urgent treatment to prevent the further neurologic problem. A design of the brain port is proposed, in which a bi-directional check valve controls the flow of the cerebrospinal fluid depending on the intracranial pressure in accordance with the other devices. Drug administration and cerebrospinal fluid drainage could be performed easily without any additional surgery other than the transplant of a brain port. The intracranial pressure value at which the cerebrospinal fluid should be drained is adjustable by altering the pressure of the drainage bag. The results of the experiment with the simulated brain system are supporting and verifying the substance of this article. | - |
dc.language | 영어 | - |
dc.language.iso | en | - |
dc.publisher | SAGE PUBLICATIONS LTD | - |
dc.relation.isPartOf | INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS | - |
dc.title | A design of brain port with a built-in bi-directional check valve improves the ventriculostomy | - |
dc.type | Article | - |
dc.type.rims | ART | - |
dc.description.journalClass | 1 | - |
dc.identifier.wosid | 000502208000001 | - |
dc.identifier.doi | 10.1177/0391398819889366 | - |
dc.identifier.bibliographicCitation | INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, v.43, no.5, pp.1 - 6 | - |
dc.description.isOpenAccess | N | - |
dc.identifier.scopusid | 2-s2.0-85077207510 | - |
dc.citation.endPage | 6 | - |
dc.citation.startPage | 1 | - |
dc.citation.title | INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS | - |
dc.citation.volume | 43 | - |
dc.citation.number | 5 | - |
dc.contributor.affiliatedAuthor | Kim, Kwang Gi | - |
dc.type.docType | Article; Early Access | - |
dc.subject.keywordAuthor | Ventriculostomy | - |
dc.subject.keywordAuthor | chemo port | - |
dc.subject.keywordAuthor | bi-directional check valve | - |
dc.subject.keywordAuthor | external ventricular drain | - |
dc.subject.keywordAuthor | intracranial pressure measurement | - |
dc.subject.keywordAuthor | cerebrospinal fluid drainage | - |
dc.relation.journalResearchArea | Engineering | - |
dc.relation.journalResearchArea | Transplantation | - |
dc.relation.journalWebOfScienceCategory | Engineering, Biomedical | - |
dc.relation.journalWebOfScienceCategory | Transplantation | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
1342, Seongnam-daero, Sujeong-gu, Seongnam-si, Gyeonggi-do, Republic of Korea(13120)031-750-5114
COPYRIGHT 2020 Gachon University All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.