Detailed Information

Cited 7 time in webofscience Cited 8 time in scopus
Metadata Downloads

Errors according to the number of registered markers used in navigation-assisted surgery of the mandible

Full metadata record
DC Field Value Language
dc.contributor.authorHwang, Young-Eun-
dc.contributor.authorKang, Sang-Hoon-
dc.contributor.authorKim, Hang-Keun-
dc.date.available2020-02-27T04:41:21Z-
dc.date.created2020-02-05-
dc.date.issued2019-02-09-
dc.identifier.issn1746-160X-
dc.identifier.urihttps://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/1819-
dc.description.abstractBackgroundThe aim of this study was to evaluate the accuracy of navigation according to the number of markers in terms of target registration errors (TREs) at each anatomical location during the registration process of the navigation system for the mandible.MethodsThe TREs were measured in five different experiments, varying only in the number of registration reference markers, which ranged from three to seven. To measure the TREs according to the number of registration reference markers, two experimental navigation devices were used: 1) Cbyon navigation surgery equipment 2) Polaris optical tracker. Both experiments were conducted to obtain the TREs at the anatomical locations of the mandible according to the number of registration markers during the navigation process. Statistical analysis was performed using the SPSS 23.0 software.ResultsAt all anatomical locations, errors were 2mm or less. Further, significant differences in the target errors measured by the Cbyon system were found according to the number of registration markers. Significant differences in the target errors measured by the Polaris optical tracker were found according to the registration markers at the posterior border only. In both groups, the target errors did not decrease as the number of registration markers increased.ConclusionsThis study demonstrates that an increase in the number of registration markers is not associated with a decrease in the TRE, and that a specific number of registration markers could reduce the TREs at each anatomical site. It is important to determine the minimum number of image registration markers at which the smallest TRE would be observed for different surgical sites.-
dc.language영어-
dc.language.isoen-
dc.publisherBMC-
dc.relation.isPartOfHEAD & FACE MEDICINE-
dc.subjectACCURACY ASSESSMENT-
dc.subjectREGISTRATION-
dc.subjectSPLINT-
dc.subjectHEAD-
dc.titleErrors according to the number of registered markers used in navigation-assisted surgery of the mandible-
dc.typeArticle-
dc.type.rimsART-
dc.description.journalClass1-
dc.identifier.wosid000458247700001-
dc.identifier.doi10.1186/s13005-019-0190-z-
dc.identifier.bibliographicCitationHEAD & FACE MEDICINE, v.15-
dc.identifier.scopusid2-s2.0-85061231742-
dc.citation.titleHEAD & FACE MEDICINE-
dc.citation.volume15-
dc.contributor.affiliatedAuthorHwang, Young-Eun-
dc.contributor.affiliatedAuthorKim, Hang-Keun-
dc.type.docTypeArticle-
dc.subject.keywordAuthorNavigation surgery-
dc.subject.keywordAuthorImage-guided surgery-
dc.subject.keywordAuthorMandible-
dc.subject.keywordAuthorRegistration-
dc.subject.keywordPlusACCURACY ASSESSMENT-
dc.subject.keywordPlusREGISTRATION-
dc.subject.keywordPlusSPLINT-
dc.subject.keywordPlusHEAD-
dc.relation.journalResearchAreaDentistry, Oral Surgery & Medicine-
dc.relation.journalWebOfScienceCategoryDentistry, Oral Surgery & Medicine-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
Files in This Item
There are no files associated with this item.
Appears in
Collections
보건과학대학 > 의용생체공학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Hang Keun photo

Kim, Hang Keun
College of IT Convergence (의공학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE