Detailed Information

Cited 2 time in webofscience Cited 2 time in scopus
Metadata Downloads

Nitinol clip distal migration and resultant popliteo-tibial artery occlusion complicating access closure by the StarClose SE vascular closure system

Authors
Choi, Dae HanKim, Myeong JinYoo, Chan JongPark, Cheol Wan
Issue Date
Feb-2017
Publisher
BMJ PUBLISHING GROUP
Keywords
Aneurysm; Complication; Coil; Device; Artery
Citation
JOURNAL OF NEUROINTERVENTIONAL SURGERY, v.9, no.2, pp.E5 - U5
Journal Title
JOURNAL OF NEUROINTERVENTIONAL SURGERY
Volume
9
Number
2
Start Page
E5
End Page
U5
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/6463
DOI
10.1136/neurintsurg-2015-012257.rep
ISSN
1759-8478
Abstract
Lower extremity ischemia following deployment of a vascular closure device for access site closure after a transfemoral endovascular procedure rarely occurs. A 68-year-old woman diagnosed with subarachnoid hemorrhage due to a ruptured anterior communicating aneurysm was treated by endovascular coil embolization. The StarClose SE device was deployed for right femoral arteriotomy closure. After 2days, critical ischemia occurred on her right lower leg due to total occlusion of the popliteo-tibial artery. Emergent surgical embolectomy was performed and the nitinol clip of the StarClose device was captured in the lumen of the tibioperoneal trunk. Although StarClose is an extravascular closure system, intravascular deployment, distal migration, and resultant critical limb ischemia can occur.
Files in This Item
There are no files associated with this item.
Appears in
Collections
의과대학 > 의학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Myeong Jin photo

Kim, Myeong Jin
College of Medicine (Department of Medicine)
Read more

Altmetrics

Total Views & Downloads

BROWSE