Fabrication of a 3D Teflon microdevice for energy free homogeneous liquid flow inside a long microchannel and its application to continuous-flow PCR
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Kieu The Loan Trinh | - |
dc.contributor.author | Wu, Wenming | - |
dc.contributor.author | Lee, Nae Yoon | - |
dc.date.available | 2020-02-27T23:41:06Z | - |
dc.date.created | 2020-02-07 | - |
dc.date.issued | 2017-02 | - |
dc.identifier.issn | 2046-2069 | - |
dc.identifier.uri | https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/7437 | - |
dc.description.abstract | We introduce a simple but robust manufacturing technology for a 3D spiral Teflon microdevice, by wrapping a polytetrafluoroethylene (PTFE) tube around a poly(dimethylsiloxane) (PDMS) mold. Semiautomated sample injection inside this fluoroplastic microdevice was achieved using one disposable syringe with flow rates systemically analyzed under different inner pressures and syringe volumes. After placing the microdevice on a single hot plate at a constant temperature of 105 C-o, a temperature gradient was formed across the 3D spiral fluoroplastic microdevice, which is suitable for the successful amplification of two DNA targets (230 bp and 409 bp) from a plasmid vector. In contrast with PDMS and poly(methyl methacrylate) (PMMA) microdevices, fewer bubbles were observed in this Teflon microdevice. The effect of the Teflon surface on amplification efficiency was also analyzed. After adding different amounts of bovine serum albumin (BSA) to the same volume of polymerase chain reaction (PCR) solution, it was found that the amplification efficiency increased when BSA concentration was adjusted from 0 to 0.24 mu g mu L-1. Futhermore, we believe that this 3D spiral fluoroplastic microdevice could be a stable platform for continuous-flow PCRs. | - |
dc.language | 영어 | - |
dc.language.iso | en | - |
dc.publisher | ROYAL SOC CHEMISTRY | - |
dc.relation.isPartOf | RSC ADVANCES | - |
dc.title | Fabrication of a 3D Teflon microdevice for energy free homogeneous liquid flow inside a long microchannel and its application to continuous-flow PCR | - |
dc.type | Article | - |
dc.type.rims | ART | - |
dc.description.journalClass | 1 | - |
dc.identifier.wosid | 000395796700002 | - |
dc.identifier.doi | 10.1039/c6ra28765d | - |
dc.identifier.bibliographicCitation | RSC ADVANCES, v.7, no.18, pp.10624 - 10630 | - |
dc.description.isOpenAccess | N | - |
dc.identifier.scopusid | 2-s2.0-85013054857 | - |
dc.citation.endPage | 10630 | - |
dc.citation.startPage | 10624 | - |
dc.citation.title | RSC ADVANCES | - |
dc.citation.volume | 7 | - |
dc.citation.number | 18 | - |
dc.contributor.affiliatedAuthor | Kieu The Loan Trinh | - |
dc.contributor.affiliatedAuthor | Wu, Wenming | - |
dc.contributor.affiliatedAuthor | Lee, Nae Yoon | - |
dc.type.docType | Article | - |
dc.subject.keywordPlus | POLYMERASE-CHAIN-REACTION | - |
dc.subject.keywordPlus | SAMPLE INJECTION | - |
dc.subject.keywordPlus | SYSTEM | - |
dc.subject.keywordPlus | CHIP | - |
dc.subject.keywordPlus | PUMP | - |
dc.subject.keywordPlus | MINIATURIZATION | - |
dc.subject.keywordPlus | MICROFLUIDICS | - |
dc.subject.keywordPlus | DELIVERY | - |
dc.subject.keywordPlus | SYRINGE | - |
dc.subject.keywordPlus | DESIGN | - |
dc.relation.journalResearchArea | Chemistry | - |
dc.relation.journalWebOfScienceCategory | Chemistry, Multidisciplinary | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
1342, Seongnam-daero, Sujeong-gu, Seongnam-si, Gyeonggi-do, Republic of Korea(13120)031-750-5114
COPYRIGHT 2020 Gachon University All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.