Detailed Information

Cited 13 time in webofscience Cited 14 time in scopus
Metadata Downloads

Porous Platinum Black-Coated Minimally Invasive Microneedles for Non-Enzymatic Continuous Glucose Monitoring in Interstitial Fluid

Full metadata record
DC Field Value Language
dc.contributor.authorChinnadayyala, S.R.-
dc.contributor.authorCho, Sungbo-
dc.date.available2021-02-16T02:40:24Z-
dc.date.created2021-01-20-
dc.date.issued2021-01-
dc.identifier.issn2079-4991-
dc.identifier.urihttps://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/79970-
dc.description.abstractIndividuals with diabetes can benefit considerably from continuous blood glucose moni-toring. To address this challenge, a proof-of-concept was performed for continuous glucose monitoring (CGM) based on an enzymeless porous nanomaterial (pNM)-modified microneedle electrode array (MNEA). The pNM sensing layer was electrochemically deposited on MNs by applying a fixed negative current of −2.5 mAcm˗2 for 400 s. The pNM-modified MNEA was packed using a biocompatible Nafion ionomer. The fabricated MNEAs were 600 × 100 × 150 µm in height, width, and thickness, respectively. The surfaces of the modified MNs were characterized by scanning electron microscopy (SEM) and energy dispersive X-ray analysis (EDX), X-ray diffraction (XRD) and X-ray photoelectron spectroscopy (XPS). The fabricated MNEAs showed a wide dynamic range (1–30 mM) in phosphate-buffered saline (PBS) and in artificial interstitial fluid (ISF), with good sensitivi-ties (PBS: 1.792 ± 0.25 µA mM−1 cm−2, ISF: 0.957 ± 0.14 µA mM−1 cm−2) and low detection limits (PBS: 7.2 µM, ISF: 22 µM). The sensor also showed high stability (loss of 3.5% at the end of 16 days), selectivity, and reproducibility (Relative standard deviations (RSD) of 1.64% and 0.70% for intra-and inter-assay, respectively) and a good response time (2 s) with great glucose recovery rates in ISF (98.7–102%). © 2020 by the authors. Licensee MDPI, Basel, Switzerland.-
dc.language영어-
dc.language.isoen-
dc.publisherMDPI-
dc.relation.isPartOfNANOMATERIALS-
dc.titlePorous Platinum Black-Coated Minimally Invasive Microneedles for Non-Enzymatic Continuous Glucose Monitoring in Interstitial Fluid-
dc.typeArticle-
dc.type.rimsART-
dc.description.journalClass1-
dc.identifier.wosid000610663100001-
dc.identifier.doi10.3390/nano11010037-
dc.identifier.bibliographicCitationNANOMATERIALS, v.11, no.1, pp.1 - 15-
dc.description.isOpenAccessN-
dc.identifier.scopusid2-s2.0-85098708014-
dc.citation.endPage15-
dc.citation.startPage1-
dc.citation.titleNANOMATERIALS-
dc.citation.volume11-
dc.citation.number1-
dc.contributor.affiliatedAuthorChinnadayyala, S.R.-
dc.contributor.affiliatedAuthorCho, Sungbo-
dc.type.docTypeArticle-
dc.subject.keywordAuthorChronoamperometry-
dc.subject.keywordAuthorCyclic voltammetry-
dc.subject.keywordAuthorElectrodeposition-
dc.subject.keywordAuthorGlucose sensor-
dc.subject.keywordAuthorMicroneedle electrode array-
dc.subject.keywordAuthorPlatinum black-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
Files in This Item
There are no files associated with this item.
Appears in
Collections
IT융합대학 > 전자공학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Somasekhar, Reddy Chinnadayyala photo

Somasekhar, Reddy Chinnadayyala
IT (Department of Electronic Engineering)
Read more

Altmetrics

Total Views & Downloads

BROWSE