Near-field clutter artifact reduction algorithm based on wavelet thresholding method in echocardiography using 3D printed cardiac phantom

Full metadata record
DC Field Value Language
dc.contributor.authorKim, Minkyoung-
dc.contributor.authorHan, Dong-Kyoon-
dc.contributor.authorLee, Youngjin-
dc.date.accessioned2022-09-21T09:40:07Z-
dc.date.available2022-09-21T09:40:07Z-
dc.date.created2022-09-15-
dc.date.issued2022-09-
dc.identifier.issn0374-4884-
dc.identifier.urihttps://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/85490-
dc.description.abstractOne of the typical sources of noise of echocardiography, near-feld clutter (NFC), is an artifact in the near feld and causes diagnostic errors with decreased accuracy. The purpose of this study is to apply an algorithm based on the wavelet threshold ing method to remove only the NFC area, without afecting the essential regions for accurate diagnosis. Ultrasound images including NFC were obtained using a self-manufactured left ventricle (LV) phantom, after which comparative evaluation was performed after applying a wavelet thresholding method-based algorithm. When the algorithm based on the wavelet thresholding method was applied to the NFC image, the root mean square error (RMSE) value decreased by 71.38%, from 35.54 to 10.17. The correlation coefcient (CC) value increased by 12.64%, from 0.87 to 0.98, and the mean structural simi larity (MSSIM) value increased by 23.68%, from 0.76 to 0.94. Finally, the universal quality index (UQI) value increased by 17.28%, from 0.81 to 0.95. In conclusion, the algorithm based on the wavelet thresholding method proved efective for removing signifcant NFCs, which afect image diagnosis in echocardiography. Furthermore, when this algorithm is used in cardiac ultrasound machines, it is expected to improve the accuracy of diagnosis by removing NFC.-
dc.language영어-
dc.language.isoen-
dc.publisher한국물리학회-
dc.relation.isPartOfJournal of the Korean Physical Society-
dc.titleNear-field clutter artifact reduction algorithm based on wavelet thresholding method in echocardiography using 3D printed cardiac phantom-
dc.typeArticle-
dc.type.rimsART-
dc.description.journalClass1-
dc.identifier.wosid000837535700003-
dc.identifier.doi10.1007/s40042-022-00512-z-
dc.identifier.bibliographicCitationJournal of the Korean Physical Society, v.81, no.5, pp.441 - 449-
dc.identifier.kciidART002875983-
dc.description.isOpenAccessN-
dc.identifier.scopusid2-s2.0-85135726982-
dc.citation.endPage449-
dc.citation.startPage441-
dc.citation.titleJournal of the Korean Physical Society-
dc.citation.volume81-
dc.citation.number5-
dc.contributor.affiliatedAuthorKim, Minkyoung-
dc.contributor.affiliatedAuthorLee, Youngjin-
dc.type.docTypeArticle-
dc.subject.keywordAuthorEchocardiography · Near-feld clutter · Artifact reduction algorithm modeling · Wavelet thresholding · 3D printing technique-
dc.relation.journalResearchAreaPhysics-
dc.relation.journalWebOfScienceCategoryPhysics, Multidisciplinary-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.description.journalRegisteredClasskci-
Files in This Item
There are no files associated with this item.
Appears in
Collections
보건과학대학 > 방사선학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Youngjin photo

Lee, Youngjin
Health Science (Dept.of Radiology)
Read more

Altmetrics

Total Views & Downloads

BROWSE