Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Simulation-Driven Triple-Tuned Array for H-1, P-31 and Na-23 Using Composite Right- and Left-Handed Transmission Line for Rat Brain at 9.4T MRI

Full metadata record
DC Field Value Language
dc.contributor.authorHernandez, Daniel-
dc.contributor.authorSeo, Minyeong-
dc.contributor.authorHan, Yeji-
dc.contributor.authorKim, Kyoung-Nam-
dc.date.accessioned2022-10-24T01:40:09Z-
dc.date.available2022-10-24T01:40:09Z-
dc.date.created2022-10-24-
dc.date.issued2022-09-
dc.identifier.issn2169-3536-
dc.identifier.urihttps://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/85841-
dc.description.abstractThe use of ultrahigh-field-strength magnetic resonance imaging (MRI), such as 9.4T, is able to acquire multi-nuclear imaging with better image quality than lower field strengths. In particular the acquisition of sodium (23Na) or phosphorus (31P) images could benefit with higher signal-to-noise ratio (SNR). The design of radiofrequency RF coils is required to achieve a uniform field and to operate at the corresponding frequency. The general method to make multiple frequency coils has the drawback of using multilayers or reducing the size of the coils, which impose restriction on the utilization of space. For the conventional multiple frequency array, the design for the coil size imposes a challenge for the optimization of SNR and field intensity. In addition, the use of multiple coils increases the coupling between each coil. To circumvent these problems, we propose the use of composite right-left handed (CRLH) transmission lines (TL), which are able to resonate to multiple frequencies. This work demonstrates a design of an array of four channels, in which each channel consists of a single CRLH element capable to resonate at three frequencies corresponding to 1H at 400 MHz, 31P at 162 MHz, and23Na at 105 MHz. The design was demonstrated with electromagnetic (EM) simulations and applied for rat brain for use in a 9.4T MRI system.-
dc.language영어-
dc.language.isoen-
dc.publisherIEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC-
dc.relation.isPartOfIEEE ACCESS-
dc.titleSimulation-Driven Triple-Tuned Array for H-1, P-31 and Na-23 Using Composite Right- and Left-Handed Transmission Line for Rat Brain at 9.4T MRI-
dc.typeArticle-
dc.type.rimsART-
dc.description.journalClass1-
dc.identifier.wosid000866425000001-
dc.identifier.doi10.1109/ACCESS.2022.3209678-
dc.identifier.bibliographicCitationIEEE ACCESS, v.10, pp.104429 - 104435-
dc.description.isOpenAccessY-
dc.identifier.scopusid2-s2.0-85139447860-
dc.citation.endPage104435-
dc.citation.startPage104429-
dc.citation.titleIEEE ACCESS-
dc.citation.volume10-
dc.contributor.affiliatedAuthorHernandez, Daniel-
dc.contributor.affiliatedAuthorSeo, Minyeong-
dc.contributor.affiliatedAuthorHan, Yeji-
dc.contributor.affiliatedAuthorKim, Kyoung-Nam-
dc.type.docTypeArticle-
dc.subject.keywordAuthorCoils-
dc.subject.keywordAuthorMagnetic resonance imaging-
dc.subject.keywordAuthorRadio access technologies-
dc.subject.keywordAuthorResonant frequency-
dc.subject.keywordAuthorPower transmission lines-
dc.subject.keywordAuthorCapacitors-
dc.subject.keywordAuthorInductors-
dc.subject.keywordAuthorComposite right-handed and left-handed (CRLH)-
dc.subject.keywordAuthormagnetic resonance imaging (MRI)-
dc.subject.keywordAuthorsmall animal imaging-
dc.subject.keywordAuthortransmission line (TL)-
dc.subject.keywordPlusCOIL-
dc.relation.journalResearchAreaComputer Science-
dc.relation.journalResearchAreaEngineering-
dc.relation.journalResearchAreaTelecommunications-
dc.relation.journalWebOfScienceCategoryComputer Science, Information Systems-
dc.relation.journalWebOfScienceCategoryEngineering, Electrical & Electronic-
dc.relation.journalWebOfScienceCategoryTelecommunications-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
Files in This Item
There are no files associated with this item.
Appears in
Collections
보건과학대학 > 의용생체공학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Han, Ye Ji photo

Han, Ye Ji
College of IT Convergence (의공학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE