Detailed Information

Cited 0 time in webofscience Cited 1 time in scopus
Metadata Downloads

Effect of surgical microscope and illuminated chopper on anterior chamber temperature

Full metadata record
DC Field Value Language
dc.contributor.authorEom, Youngsub-
dc.contributor.authorLee, Young Joo-
dc.contributor.authorSong, Jong Suk-
dc.contributor.authorKim, Hyo Myung-
dc.contributor.authorNam, Dong Heun-
dc.date.accessioned2023-03-14T07:40:34Z-
dc.date.available2023-03-14T07:40:34Z-
dc.date.created2023-03-13-
dc.date.issued2023-01-
dc.identifier.issn1471-2415-
dc.identifier.urihttps://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/87112-
dc.description.abstractBackgroundTo evaluate the effect of the light intensity of the surgical microscope and illuminated chopper on the anterior chamber temperature.Study designExperimental study.MethodsA model eye (Kitaro WetLab System; FCI Ophthalmics, Pembroke, MA, USA) was used in this experimental study. The illuminance of a surgical microscope (Leica M300; Leica Microsystems, Wetzlar, Germany) and illuminated chopper (iChopper NAM-25 GB; Oculight, Korea) with a light source (iVision; Oculight) was measured using an illuminometer. In addition, the temperature in the anterior chamber of the model eye filled with balanced salt solution when using the surgical microscope with a light intensity from level 1 to level 6 and the illuminated chopper at 99% light intensity was measured for 10 min.ResultsThe anterior chamber temperature was increased by 0.2, 0.5, 1.0, and 1.4 celcius when using the surgical microscope at level 3 (10050 lux), 4 (16490 lux), 5 (24900 lux), and 6 (32500 lux), respectively, for 10 min. The illuminated chopper at 99% light intensity (14893 lux) positioned in the anterior chamber increased the anterior chamber temperature by 0.2 degrees C after 10 min, which was equal to the increase in the temperature caused by the surgical microscope at level 3.ConclusionThe photothermal effect of the illuminated chopper directly positioned in the anterior chamber appeared to be similar to that of a microscope with similar illuminance. Therefore, the illuminated chopper is safe in terms of anterior chamber temperature changes in cataract surgery.-
dc.language영어-
dc.language.isoen-
dc.publisherBMC-
dc.relation.isPartOfBMC OPHTHALMOLOGY-
dc.titleEffect of surgical microscope and illuminated chopper on anterior chamber temperature-
dc.typeArticle-
dc.type.rimsART-
dc.description.journalClass1-
dc.identifier.wosid000920135400005-
dc.identifier.doi10.1186/s12886-023-02784-w-
dc.identifier.bibliographicCitationBMC OPHTHALMOLOGY, v.23, no.1-
dc.description.isOpenAccessY-
dc.identifier.scopusid2-s2.0-85146784998-
dc.citation.titleBMC OPHTHALMOLOGY-
dc.citation.volume23-
dc.citation.number1-
dc.contributor.affiliatedAuthorNam, Dong Heun-
dc.type.docTypeArticle-
dc.subject.keywordAuthorMicroscope-
dc.subject.keywordAuthorIlluminated chopper-
dc.subject.keywordAuthorTemperature-
dc.subject.keywordPlusINTRACAMERAL ILLUMINATION-
dc.subject.keywordPlusCATARACT-SURGERY-
dc.subject.keywordPlusPHOTOTOXICITY-
dc.subject.keywordPlusENDOILLUMINATION-
dc.subject.keywordPlusPARAMETERS-
dc.relation.journalResearchAreaOphthalmology-
dc.relation.journalWebOfScienceCategoryOphthalmology-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
Files in This Item
There are no files associated with this item.
Appears in
Collections
의과대학 > 의학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Nam, Dong Heun photo

Nam, Dong Heun
College of Medicine (Department of Medicine)
Read more

Altmetrics

Total Views & Downloads

BROWSE