심장 전기생리학 검사 시 조건 변화에 따른 환자 피폭 선량 평가Evaluation of Patient Exposure Dose during Cardiac Electrophysiology Study under Various Conditions

Other Titles
Evaluation of Patient Exposure Dose during Cardiac Electrophysiology Study under Various Conditions
Authors
고성빈안성민
Issue Date
Dec-2023
Publisher
대한방사선과학회
Keywords
Electrophysiology Study; Fluoroscopy; FPS; Fluoflavor; Dose Reduction; 심장 전기생리학 검사; 연속 투시촬영; 초당 프레임 수; Fluoflavor; 피폭선량 감소
Citation
방사선기술과학, v.46, no.6, pp 501 - 508
Pages
8
Journal Title
방사선기술과학
Volume
46
Number
6
Start Page
501
End Page
508
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/89896
DOI
10.17946/JRST.2023.46.6.501
ISSN
2288-3509
2384-1168
Abstract
This study used a adult absorption dose phantom (CIRS model 701-G, USA) made of human equivalent material and the vascular imaging equipment Allura Xper FD 20 (Philips, Netherlands). Optically stimulated luminescent dosimeters (OSLD) were inserted into the anatomical positions corresponding to each organ, and the exposure dose was measured. Dose area product (DAP) and air kerma (AK) measured by the dose meter in the equipment were compared. Continuous imaging was performed at two angles for a total of 20 minutes, with a frame per seconds of 3.75 and 7.5 fps and an FOV of 42cm, 37cm, and 31cm, respectively, under the conditions of fluoflavor I, II, and III, each selected for 5 repetitions. This study was found that selecting a lower fps was the most effective way to reduce patient exposure dose, and adjusting the fluoflavor was a good alternative method for reducing patient exposure dose at high fps. Therefore the method of condition change with the greatest dose reduction effect is to set the minimum FPS and can reduce patient exposure dose according to geometric conditions and fluoflavor characteristics.
Files in This Item
There are no files associated with this item.
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Ahn, Sung Min photo

Ahn, Sung Min
Health Science (Dept.of Radiology)
Read more

Altmetrics

Total Views & Downloads

BROWSE