Light exposure from microscope versus intracameral illumination during cataract surgery

Full metadata record
DC Field Value Language
dc.contributor.authorKim, Yu Jeong-
dc.contributor.authorNam, Dong Heun-
dc.contributor.authorKim, Young Jae-
dc.contributor.authorKim, Kwang Gi-
dc.contributor.authorKim, Seong-Woo-
dc.contributor.authorChung, Tae-Young-
dc.contributor.authorLee, Sung Jin-
dc.contributor.authorPark, Kyu Hyung-
dc.date.available2020-02-27T02:21:49Z-
dc.date.created2020-02-04-
dc.date.issued2019-10-
dc.identifier.issn0301-4738-
dc.identifier.urihttps://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/953-
dc.description.abstractPurpose: To evaluate light exposure from microscope versus intracameral illuminations to patient's and surgeon's retina during cataract surgery. Methods: Thirty consecutive patients who had cataract surgery using microscope and intracameral illuminations. At the point of the ocular of an operating microscope, optical illuminance and irradiance from the microscope illumination (60, 40, 20% intensity) and the intracameral illumination (60% intensity) were measured using a light meter and a spectrometer at a pause after lens capsule polishing in cataract surgery. Results: Average illuminance (lux) was 1.46, 0.66, 0.27, and 0.1 from 60%, 40%, 20% intensity microscope illuminations and 60% intracameral illumination. Average total spectral irradiance (mu W/cm2) was 1.25, 0.65, 0.26, and 0.03 from 60%, 40%, 20% intensity microscope illuminations and 60% intracameral illumination. Conclusion: Microscope ocular illuminance and irradiance during cataract surgery were higher in the microscope illumination than in the intracameral illumination. It suggests that light exposure reaching patient's and surgeon's retina during cataract surgery is lower in the intracameral illumination than in the microscope illumination.-
dc.language영어-
dc.language.isoen-
dc.publisherWOLTERS KLUWER MEDKNOW PUBLICATIONS-
dc.relation.isPartOfINDIAN JOURNAL OF OPHTHALMOLOGY-
dc.titleLight exposure from microscope versus intracameral illumination during cataract surgery-
dc.typeArticle-
dc.type.rimsART-
dc.description.journalClass1-
dc.identifier.wosid000488008300028-
dc.identifier.doi10.4103/ijo.IJO_316_19-
dc.identifier.bibliographicCitationINDIAN JOURNAL OF OPHTHALMOLOGY, v.67, no.10, pp.1624 - +-
dc.identifier.scopusid2-s2.0-85072571870-
dc.citation.endPage+-
dc.citation.startPage1624-
dc.citation.titleINDIAN JOURNAL OF OPHTHALMOLOGY-
dc.citation.volume67-
dc.citation.number10-
dc.contributor.affiliatedAuthorNam, Dong Heun-
dc.contributor.affiliatedAuthorKim, Young Jae-
dc.contributor.affiliatedAuthorKim, Kwang Gi-
dc.type.docTypeArticle-
dc.subject.keywordAuthorCataract surgery-
dc.subject.keywordAuthorilluminance-
dc.subject.keywordAuthorintracameral illumination-
dc.subject.keywordAuthorirradiance-
dc.subject.keywordAuthorlight exposure-
dc.subject.keywordAuthorphototoxicity-
dc.subject.keywordPlusOPERATING MICROSCOPE-
dc.subject.keywordPlusEYES-
dc.subject.keywordPlusCAPSULE-
dc.subject.keywordPlusSAFE-
dc.relation.journalResearchAreaOphthalmology-
dc.relation.journalWebOfScienceCategoryOphthalmology-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
Files in This Item
There are no files associated with this item.
Appears in
Collections
보건과학대학 > 의용생체공학과 > 1. Journal Articles
의과대학 > 의학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Kwang Gi photo

Kim, Kwang Gi
College of IT Convergence (의공학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE