Detailed Information

Cited 3 time in webofscience Cited 0 time in scopus
Metadata Downloads

Adjustable RF Transmitter Head Coil: Improving Transmit Efficiency With SAR Management for 7-T Magnetic Resonance Imaging

Full metadata record
DC Field Value Language
dc.contributor.authorCho, Youngdae-
dc.contributor.authorBasir, Abdul-
dc.contributor.authorYoo, Hyoung suk-
dc.date.accessioned2022-07-06T20:36:26Z-
dc.date.available2022-07-06T20:36:26Z-
dc.date.created2021-07-14-
dc.date.issued2021-05-
dc.identifier.issn0018-9480-
dc.identifier.urihttps://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/141978-
dc.description.abstractIn ultrahigh field magnetic resonance imaging (MRI) using a static magnetic field over 7 T, the inhomogeneity in the time-varying magnetic field ( B1), which is caused by its short wavelength, is mitigated by using a multichannel radio frequency (RF) coil as an antenna. However, the conventional multichannel RF coil has a fixed configuration; consequently, its transmit efficiency (TE) changes according to the head shape of the patient. In this study, a geometrically adjustable eight-channel RF transmitter head coil for 7-T MRI was proposed. All resonators and supporting structures (called bridges) in the coil have interlocking teeth at their sides, allowing each channel to adjust its geometric location according to the head shape of the patient. Scenarios for conventional and proposed coils in three different configurations were simulated and compared in terms of the transmitted component of the B1 field ( B+ 1 field) and specific absorption rate (SAR) maps. The simulation results indicate that, for a given configuration and head shape, the TE of the proposed coil was higher than that of the conventional coil, while the SAR efficiency remained unchanged. Furthermore, the S-parameter matrix of the proposed coil and the measured B1 field produced by the two channels facing each other show that the proposed coil resonates at 298 MHz with acceptable mutual coupling and effectively improves the B1 field intensity. Moreover, the obtained signal-to-noise ratio (SNR) maps and magnetic resonance (MR) images prove that the proposed coil is an efficient candidate for an RF transmitter or transceiver head coil used in 7-T MRI.-
dc.language영어-
dc.language.isoen-
dc.publisherIEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC-
dc.titleAdjustable RF Transmitter Head Coil: Improving Transmit Efficiency With SAR Management for 7-T Magnetic Resonance Imaging-
dc.typeArticle-
dc.contributor.affiliatedAuthorYoo, Hyoung suk-
dc.identifier.doi10.1109/TMTT.2021.3057620-
dc.identifier.scopusid2-s2.0-85101779390-
dc.identifier.wosid000662675000023-
dc.identifier.bibliographicCitationIEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, v.69, no.5, pp.2686 - 2696-
dc.relation.isPartOfIEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES-
dc.citation.titleIEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES-
dc.citation.volume69-
dc.citation.number5-
dc.citation.startPage2686-
dc.citation.endPage2696-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.isOpenAccessN-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaEngineering-
dc.relation.journalWebOfScienceCategoryEngineering, Electrical & Electronic-
dc.subject.keywordPlusSTEPPED IMPEDANCE RESONATORS-
dc.subject.keywordPlus7 T-
dc.subject.keywordPlusPHASED-ARRAY-
dc.subject.keywordPlusDESIGN-
dc.subject.keywordPlusMRI-
dc.subject.keywordPlusRECEIVE-
dc.subject.keywordPlusFIELDS-
dc.subject.keywordAuthorRadio frequency-
dc.subject.keywordAuthorHead-
dc.subject.keywordAuthorMagnetic resonance imaging-
dc.subject.keywordAuthorShape-
dc.subject.keywordAuthorMagnetic heads-
dc.subject.keywordAuthorResonators-
dc.subject.keywordAuthorShape measurement-
dc.subject.keywordAuthorAdjustable coil-
dc.subject.keywordAuthormagnetic resonance imaging (MRI)-
dc.subject.keywordAuthorradio frequency (RF) coil-
dc.subject.keywordAuthor7 T-
dc.subject.keywordAuthortransmitter head coil-
dc.identifier.urlhttps://ieeexplore.ieee.org/document/9358238-
Files in This Item
Go to Link
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoo, Hyoungsuk photo

Yoo, Hyoungsuk
COLLEGE OF ENGINEERING (서울 바이오메디컬공학전공)
Read more

Altmetrics

Total Views & Downloads

BROWSE