Detailed Information

Cited 16 time in webofscience Cited 16 time in scopus
Metadata Downloads

Radiative Near-Field Wireless Power Transfer to Scalp-Implantable Biotelemetric Device

Authors
Shah, Syed Ahson AliYoo, Hyoungsuk
Issue Date
Jul-2020
Publisher
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
Keywords
Antenna measurements; Monitoring; Radio frequency; Antennas; Telemetry; Implants; Biomedical measurement; Dual band; energy conversion; Fresnel zone; implant; intracranial pressure (ICP); link budget; radio frequency (RF)-to-dc rectifier
Citation
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, v.68, no.7, pp.2944 - 2953
Indexed
SCIE
SCOPUS
Journal Title
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES
Volume
68
Number
7
Start Page
2944
End Page
2953
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/145425
DOI
10.1109/TMTT.2020.2985356
ISSN
0018-9480
Abstract
Monitoring of elevated intracranial pressure (ICP) is a lifesaving procedure. This article presents an ultrasmall batteryless implantable system for ICP monitoring comprising an off-body power transmitter (Tx) and in-body biotelemetric receiver (Rx). Due to the small implantable antenna (5.6 mm x 6 mm x 0.2 mm = 6.72 mm(3)), the device exhibits dual-band characteristics (i.e., 915 and 1900 MHz) for simultaneous power transmission and data telemetry. The implant is powered wirelessly in the radiative near field (1900 MHz) for enhanced power transfer efficiency (PTE). Moreover, the structure demonstrates the measured peak gain values of -26.8 and -18.8 dBi with the impedance-matched bandwidths of 9.83% and 27.9% at 915 and 1900 MHz, respectively. The wireless PTE of the device was also analyzed in terms of distance variations, and a maximum PTE up to -25.9 dB at 20 mm (0.1267 lambda) Tx-Rx separation was achieved. The rectifier attained a maximum power conversion efficiency of 82% at 2-dBm input power. Simulations using the finite-element method and finite difference time domain were performed to evaluate the biotelemetric implantable system. For validation, measurements were conducted in a saline-filled human head phantom, as well as in minced pork. The biotelemetric system yields good agreement between the measured and simulation results.
Files in This Item
Go to Link
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoo, Hyoungsuk photo

Yoo, Hyoungsuk
COLLEGE OF ENGINEERING (서울 바이오메디컬공학전공)
Read more

Altmetrics

Total Views & Downloads

BROWSE