Efficient Wireless Power Transfer System With a Miniaturized Quad-Band Implantable Antenna for Deep-Body Multitasking Implants

Full metadata record
DC Field Value Language
dc.contributor.authorBasir, Abdul-
dc.contributor.authorYoo, Hyoungsuk-
dc.date.accessioned2022-07-08T03:31:30Z-
dc.date.available2022-07-08T03:31:30Z-
dc.date.created2021-05-12-
dc.date.issued2020-05-
dc.identifier.issn0018-9480-
dc.identifier.urihttps://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/145741-
dc.description.abstractPassive operation and battery-charging of deep-body implants can be insured through wireless power transfer (WPT) technologies. However, the power transfer efficiency (PTE) is constrained by device miniaturization and implantation depth. This study proposes a complete WPT system consisting of a patterned WPT transmitter (Tx), an efficient rectifier, and an antenna integrated with the system. The WPT Tx had a size of 6 cm 6 cm and was optimized to focus the power on the deep-tissue implants at 1470 MHz. The voltage doubler was optimized at 1470 MHz, had a small size of 5 mm 10 mm, and exhibited a high radio frequency (RF)-to-direct current (dc) conversion efficiency of 90% at 2-dBm RF input power. Moreover, the implantable antenna occupies a small volume of 8.43 mm(3) and supports quad-band operations: telemetry at 403 and 915 MHz, WPT at the midfield band of 1470 MHz, and control signaling at 2.4 GHz. First, the fabricated prototypes were measured individually in minced pork, in the American Society for Testing and Materials (ASTM) model, and in the saline-filled 3-D head phantom. While operating collectively as an integrated system, the PTE of the system was measured. Additionally, to enhance the PTE of the WPT system, a high-dielectric matching layer = 78 was used between the WPT Tx and the phantom. Furthermore, to demonstrate the PTE of the WPT system, the voltage doubler was integrated with the implantable antenna, encapsulated in a 3-D-printed capsule endoscope, and its PTE was measured in a saline solution and minced pork. Finally, the compliance of the WPT system with the human safety standards was analyzed and found that the system solely satisfied the safety limits. It is evident from the experimental results that the system can transfer 6.7-mW power to millimeter-sized implants located 5-cm deep in tissues.-
dc.language영어-
dc.language.isoen-
dc.publisherIEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC-
dc.titleEfficient Wireless Power Transfer System With a Miniaturized Quad-Band Implantable Antenna for Deep-Body Multitasking Implants-
dc.typeArticle-
dc.contributor.affiliatedAuthorYoo, Hyoungsuk-
dc.identifier.doi10.1109/TMTT.2020.2965938-
dc.identifier.scopusid2-s2.0-85084855972-
dc.identifier.wosid000534244700032-
dc.identifier.bibliographicCitationIEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, v.68, no.5, pp.1943 - 1953-
dc.relation.isPartOfIEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES-
dc.citation.titleIEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES-
dc.citation.volume68-
dc.citation.number5-
dc.citation.startPage1943-
dc.citation.endPage1953-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.isOpenAccessN-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaEngineering-
dc.relation.journalWebOfScienceCategoryEngineering, Electrical & Electronic-
dc.subject.keywordPlusRESISTANCE COMPRESSION NETWORKS-
dc.subject.keywordPlusDESIGN-
dc.subject.keywordPlusTISSUE-
dc.subject.keywordAuthorBatteryless implants-
dc.subject.keywordAuthorcapsule endoscopy-
dc.subject.keywordAuthorenergy harvesting-
dc.subject.keywordAuthorpacemaker-
dc.subject.keywordAuthorpower transfer efficiency (PTE)-
dc.identifier.urlhttps://ieeexplore.ieee.org/document/8978635-
Files in This Item
Go to Link
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoo, Hyoungsuk photo

Yoo, Hyoungsuk
COLLEGE OF ENGINEERING (서울 바이오메디컬공학전공)
Read more

Altmetrics

Total Views & Downloads

BROWSE