dc.description.abstractPhotoplethysmography (PPG) is an easy and convenient method by which to measure heart rate (HR). However, PPG signals that optically measure volumetric changes in blood are not robust to motion artifacts. In this paper, we develop a PPG measuring system based on multi-channel sensors with multiple wavelengths and propose a motion artifact reduction algorithm using independent component analysis (ICA). We also propose a truncated singular value decomposition for 12-channel PPG signals, which contain direction and depth information measured using the developed multi-channel PPG measurement system. The performance of the proposed method is evaluated against the R-peaks of an electrocardiogram in terms of sensitivity (Se), positive predictive value (PPV), and failed detection rate (FDR). The experimental results show that Se, PPV, and FDR were 99%, 99.55%, and 0.45% for walking, 96.28%, 99.24%, and 0.77% for fast walking, and 82.49%, 99.83%, and 0.17% for running, respectively. The evaluation shows that the proposed method is effective in reducing errors in HR estimation from PPG signals with motion artifacts in intensive motion situations such as fast walking and running.-
dc.language영어-
dc.language.isoen-
dc.publisherMDPI-
dc.titleMotion Artifact Reduction in Wearable Photoplethysmography Based on Multi-Channel Sensors with Multiple Wavelengths-
dc.typeArticle-
dc.contributor.affiliatedAuthorPark, Hoon-Ki-
dc.contributor.affiliatedAuthorKim, In Young-
dc.identifier.doi10.3390/s20051493-
dc.identifier.scopusid2-s2.0-85081129242-
dc.identifier.wosid000525271500255-
dc.identifier.bibliographicCitationSENSORS, v.20, no.5, pp.1 - 14-
dc.relation.isPartOfSENSORS-
dc.citation.titleSENSORS-
dc.citation.volume20-
dc.citation.number5-
dc.citation.startPage1-
dc.citation.endPage14-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.isOpenAccessY-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaChemistry-
dc.relation.journalResearchAreaEngineering-
dc.relation.journalResearchAreaInstruments & Instrumentation-
dc.relation.journalWebOfScienceCategoryChemistry, Analytical-
dc.relation.journalWebOfScienceCategoryEngineering, Electrical & Electronic-
dc.relation.journalWebOfScienceCategoryInstruments & Instrumentation-
dc.subject.keywordPlusIndependent component analysis-
dc.subject.keywordPlusMotion sensors-
dc.subject.keywordPlusPhotoplethysmography-
dc.subject.keywordPlusSingular value decomposition-
dc.subject.keywordPlusWearable sensors-
dc.subject.keywordPlusIndependent component analysis-
dc.subject.keywordPlusMotion sensors-
dc.subject.keywordPlusPhotoplethysmography-
dc.subject.keywordPlusSingular value decomposition-
dc.subject.keywordPlusIndependent component analysis(ICA)-
dc.subject.keywordPlusMotion artifact-
dc.subject.keywordPlusMotion artifact reduction-
dc.subject.keywordPlusMulti channel sensors-
dc.subject.keywordPlusMulti-wavelengths-
dc.subject.keywordPlusPhotoplethysmography (PPG)-
dc.subject.keywordPlusPositive predictive values-
dc.subject.keywordPlusTruncated singular value decomposition-
dc.subject.keywordPlusadult-
dc.subject.keywordPlusartifact-
dc.subject.keywordPlusdevices-
dc.subject.keywordPluselectrocardiography-
dc.subject.keywordPluselectronic device-
dc.subject.keywordPlusequipment design-
dc.subject.keywordPlusfemale-
dc.subject.keywordPlusheart rate-
dc.subject.keywordPlushuman-
dc.subject.keywordPlusmale-
dc.subject.keywordPlusphotoelectric plethysmography-
dc.subject.keywordPlusphysiology-
dc.subject.keywordPlussignal processing-
dc.subject.keywordPluswalking-
dc.subject.keywordPlusyoung adult-
dc.subject.keywordAuthorphotoplethysmography-
dc.subject.keywordAuthormotion artifact-
dc.subject.keywordAuthorindependent component analysis-
dc.subject.keywordAuthormulti-wavelength-
dc.identifier.urlhttps://www.mdpi.com/1424-8220/20/5/1493-
Files in This Item
Appears in
Collections
서울 의과대학 > 서울 의공학교실 > 1. Journal Articles
서울 의과대학 > 서울 가정의학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, In Young photo

Kim, In Young
COLLEGE OF MEDICINE (DEPARTMENT OF BIOMEDICAL ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE