Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

The Use of Catheter Mount Will Result in More Reliable Carbon Dioxide Monitoring under Fluid Exposing Conditions

Full metadata record
DC Field Value Language
dc.contributor.authorCho, Yongil-
dc.contributor.authorKim, Wonhee-
dc.contributor.authorLim, Tae Ho-
dc.contributor.authorChoi, Hyuk Joong-
dc.contributor.authorOh, Jaehoon-
dc.contributor.authorKang, Bossng-
dc.contributor.authorKim, Youjin-
dc.contributor.authorKim, In Young-
dc.date.accessioned2022-07-09T13:55:34Z-
dc.date.available2022-07-09T13:55:34Z-
dc.date.created2021-05-11-
dc.date.issued2019-07-
dc.identifier.issn2090-2840-
dc.identifier.urihttps://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/147505-
dc.description.abstractIntroduction. Capnometer can be readily malfunctioned by fluid exposure during treatment of critically ill patients. This study aimed to determine whether placing capnometer distant from the endotracheal tube by connecting direct connect catheter mount (DCCM) is effective in yielding reliable end-tidal carbon dioxide (ETCO2) by reducing capnometer malfunctioning caused by water exposure. Methods. In 25 healthy adults, a prospective, open label, crossover study was conducted to examine the effect of DCCM in mainstream and microstream capnometers under water exposing conditions. The primary endpoint was the comparison of ETCO2 between proximal DCCM (pDCCM) and distal DCCM (dDCCM). Results. For mainstream capnometers, mean ETCO2 was significantly (p < 0.001) higher in dDCCM compared to pDCCM under water exposing conditions (29.5 vs. 19.0 with 5 ml; 33.8 vs. 21.2 with 10 ml; mmHg). Likewise, for microstream capnometers, ETCO2 was greatly higher (p < 0.001) in dDCCM compared to pDCCM (30.5 vs. 13.9 with 5 ml; 29.9 vs.11.4 with 10 mL; mmHg). ETCO2 measured by dDCCM was reliable in microstream settings, whereas it was unreliable in mainstream (correlation coefficient 0.88 vs. 0.27). Conclusions. Application of DCCM onto the capnometer setting seems to be effective in reducing capnometer malfunctioning under fluid exposing conditions, which is obvious in microstream capnometer by producing more reliable ETCO2.-
dc.language영어-
dc.language.isoen-
dc.publisherHINDAWI LTD-
dc.titleThe Use of Catheter Mount Will Result in More Reliable Carbon Dioxide Monitoring under Fluid Exposing Conditions-
dc.typeArticle-
dc.contributor.affiliatedAuthorCho, Yongil-
dc.contributor.affiliatedAuthorLim, Tae Ho-
dc.contributor.affiliatedAuthorChoi, Hyuk Joong-
dc.contributor.affiliatedAuthorOh, Jaehoon-
dc.identifier.doi10.1155/2019/4120127-
dc.identifier.wosid000475584200001-
dc.identifier.bibliographicCitationEMERGENCY MEDICINE INTERNATIONAL, v.2019, pp.1 - 8-
dc.relation.isPartOfEMERGENCY MEDICINE INTERNATIONAL-
dc.citation.titleEMERGENCY MEDICINE INTERNATIONAL-
dc.citation.volume2019-
dc.citation.startPage1-
dc.citation.endPage8-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.isOpenAccessY-
dc.description.journalRegisteredClassscie-
dc.relation.journalResearchAreaEmergency Medicine-
dc.relation.journalWebOfScienceCategoryEmergency Medicine-
dc.subject.keywordPlusRESUSCITATION-
dc.subject.keywordPlusGUIDELINES-
dc.identifier.urlhttps://www.hindawi.com/journals/emi/2019/4120127/-
Files in This Item
Appears in
Collections
서울 의과대학 > 서울 응급의학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lim, Tae Ho photo

Lim, Tae Ho
COLLEGE OF MEDICINE (DEPARTMENT OF EMERGENCY MEDICINE)
Read more

Altmetrics

Total Views & Downloads

BROWSE