Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Effect of a new heated and humidified breathing circuit with a fluid-warming device on intraoperative core temperature: a prospective randomized study

Authors
Kim, Eu geneLee, Sue-YoungLim, Young-JinChoi, Jung-YoonJeon, Young-TaeHwang, Jung-WonPark, Hee-Pyoung
Issue Date
Aug-2015
Publisher
SPRINGER JAPAN KK
Keywords
Breathing circuit; Infusion fluid temperature; Intraoperative core temperature
Citation
JOURNAL OF ANESTHESIA, v.29, no.4, pp.499 - 507
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF ANESTHESIA
Volume
29
Number
4
Start Page
499
End Page
507
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/156562
DOI
10.1007/s00540-015-1994-z
ISSN
0913-8668
Abstract
The effect of the Mega Acer kit(@), a new heated and humidified breathing circuit (HHBC) containing a fluid-warming device, was investigated on intraoperative core temperature (T (c)). A total of 102 patients undergoing elective craniotomies were randomly divided into three groups based on the breathing circuit used: a conventional breathing circuit (group C, n = 34), a Fisher & Paykel HHBC (group F, n = 34), and the Mega (group M, n = 34). From baseline to the end of the surgery, T (c) and infusion fluid temperature (T (f)) were recorded at 15-min intervals. If T (c) became lower than 35.5 A degrees C, a forced-air warmer was used. Baseline temperatures were 36.7 +/- A 0.3, 36.6 +/- A 0.2, and 36.6 +/- A 0.2 A degrees C in groups C, M, and F, respectively. T (c) at the end of surgery dropped from baseline values by 1.0 +/- A 0.4, 0.5 +/- A 0.5, and 0.8 +/- A 0.5 A degrees C in groups C, M, and F, respectively. From 60 min of post-induction to the end of surgery, T (c) was higher in group M than group C (p < 0.05). From 105 min of post-induction to the end of surgery, T (c) was higher in group M than group F (p < 0.05). The number of patients receiving forced-air warmer and total forced-air warmer using time were significantly lower in group M than groups C and F (p < 0.05). T (f) was higher in group M than groups C and F throughout the study period (31.0 +/- A 1.0 vs. 23.5 +/- A 0.5 and 24.0 +/- A 0.4 A degrees C; p < 0.01). The Mega significantly reduced the drop in intraoperative T (c) by delivering warm fluids, compared with the other breathing circuits tested.
Files in This Item
Go to Link
Appears in
Collections
서울 의과대학 > 서울 마취통증의학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Eugene photo

Kim, Eugene
COLLEGE OF MEDICINE (DEPARTMENT OF ANESTHESIA AND MEDICINE)
Read more

Altmetrics

Total Views & Downloads

BROWSE