2D Sparse Array Transducer Optimization for 3D Ultrasound Imaging
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | Choi, Jae Hoon | - |
| dc.contributor.author | Park, Kwan Kyu | - |
| dc.date.accessioned | 2022-07-16T01:30:19Z | - |
| dc.date.available | 2022-07-16T01:30:19Z | - |
| dc.date.issued | 2014-12 | - |
| dc.identifier.issn | 1225-7842 | - |
| dc.identifier.issn | 2287-402X | - |
| dc.identifier.uri | https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/158404 | - |
| dc.description.abstract | A 3D ultrasound image is desired in many medical examinations. However, the implementation of a 2D array, which is needed for a 3D image, is challenging with respect to fabrication, interconnection and cabling. A 2D sparse array, which needs fewer elements than a dense array, is a realistic way to achieve 3D images. Because the number of ways the elements can be placed in an array is extremely large, a method for optimizing the array configuration is needed. Previous research placed the target point far from the transducer array, making it impossible to optimize the array in the operating range. In our study, we focused on optimizing a 2D sparse array transducer for 3D imaging by using a simulated annealing method. We compared the far-field optimization method with the near-field optimization method by analyzing a point-spread function(PSF). The resolution of the optimized sparse array is comparable to that of the dense array. | - |
| dc.format.extent | 6 | - |
| dc.language | 영어 | - |
| dc.language.iso | ENG | - |
| dc.publisher | 한국비파괴검사학회 | - |
| dc.title | 2D Sparse Array Transducer Optimization for 3D Ultrasound Imaging | - |
| dc.type | Article | - |
| dc.publisher.location | 대한민국 | - |
| dc.identifier.doi | 10.7779/JKSNT.2014.34.6.441 | - |
| dc.identifier.bibliographicCitation | 비파괴검사학회지, v.34, no.6, pp 441 - 446 | - |
| dc.citation.title | 비파괴검사학회지 | - |
| dc.citation.volume | 34 | - |
| dc.citation.number | 6 | - |
| dc.citation.startPage | 441 | - |
| dc.citation.endPage | 446 | - |
| dc.identifier.kciid | ART001945915 | - |
| dc.description.isOpenAccess | N | - |
| dc.description.journalRegisteredClass | kci | - |
| dc.subject.keywordAuthor | Sparse Array Transducer | - |
| dc.subject.keywordAuthor | Simulated Annealing | - |
| dc.subject.keywordAuthor | Optimization | - |
| dc.subject.keywordAuthor | Ultrasound 3D Imaging | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
222, Wangsimni-ro, Seongdong-gu, Seoul, 04763, Korea+82-2-2220-1366
COPYRIGHT © 2024 HANYANG UNIVERSITY.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.