Does the accuracy of blood pressure measurement correlate with hearing loss of the observer?
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Song, Soohwa | - |
dc.contributor.author | Lee, Jongshill | - |
dc.contributor.author | Chee, Youngjoon | - |
dc.contributor.author | Jang, Dong Pyo | - |
dc.contributor.author | Kim, In Young | - |
dc.date.accessioned | 2022-07-16T06:10:21Z | - |
dc.date.available | 2022-07-16T06:10:21Z | - |
dc.date.created | 2021-05-12 | - |
dc.date.issued | 2014-02 | - |
dc.identifier.issn | 1359-5237 | - |
dc.identifier.uri | https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/160735 | - |
dc.description.abstract | ObjectiveThe auscultatory method is influenced by the hearing level of the observers. If the observer has hearing loss, it is possible to measure blood pressure inaccurately by misreading the Korotkoff sounds at systolic blood pressure (SBP) and diastolic blood pressure (DBP). Because of the potential clinical problems this discrepancy may cause, we used a hearing loss simulator to determine how hearing level affects the accuracy of blood pressure measurements.Materials and methodsTwo data sets (data set A, 32 Korotkoff sound video clips recorded by the British Hypertension Society; data set B, 28 Korotkoff sound data acquired from the Korotkoff sound recording system developed by Hanyang University) were used and all the data were attenuated to simulate a hearing loss of 5, 10, 15, 20, and 25 dB using the hearing loss simulator. Five observers with normal hearing assessed the blood pressures from these data sets and the differences between the values measured from the original recordings (no attenuation) and the attenuated versions were analyzed.ResultsGreater attenuation of the Korotkoff sounds, or greater hearing loss, resulted in larger blood pressure measurement differences when compared with the original data. When measuring blood pressure with hearing loss, the SBP tended to be underestimated and the DBP was overestimated. The mean differences between the original data and the 25 dB hearing loss data for the two data sets combined were 1.552.71 and -4.32 +/- 4.21 mmHg for SBP and DBP, respectively.ConclusionThis experiment showed that the accuracy of blood pressure measurements using the auscultatory method is affected by observer hearing level. Therefore, to reduce possible error using the auscultatory method, observers' hearing should be tested. | - |
dc.language | 영어 | - |
dc.language.iso | en | - |
dc.publisher | LIPPINCOTT WILLIAMS & WILKINS | - |
dc.title | Does the accuracy of blood pressure measurement correlate with hearing loss of the observer? | - |
dc.type | Article | - |
dc.contributor.affiliatedAuthor | Jang, Dong Pyo | - |
dc.contributor.affiliatedAuthor | Kim, In Young | - |
dc.identifier.doi | 10.1097/MBP.0000000000000016 | - |
dc.identifier.scopusid | 2-s2.0-84894070194 | - |
dc.identifier.wosid | 000330880200003 | - |
dc.identifier.bibliographicCitation | BLOOD PRESSURE MONITORING, v.19, no.1, pp.14 - 18 | - |
dc.relation.isPartOf | BLOOD PRESSURE MONITORING | - |
dc.citation.title | BLOOD PRESSURE MONITORING | - |
dc.citation.volume | 19 | - |
dc.citation.number | 1 | - |
dc.citation.startPage | 14 | - |
dc.citation.endPage | 18 | - |
dc.type.rims | ART | - |
dc.type.docType | Article | - |
dc.description.journalClass | 1 | - |
dc.description.isOpenAccess | N | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
dc.relation.journalResearchArea | Cardiovascular System & Cardiology | - |
dc.relation.journalWebOfScienceCategory | Peripheral Vascular Disease | - |
dc.subject.keywordPlus | LOUDNESS RECRUITMENT | - |
dc.subject.keywordPlus | THRESHOLD ELEVATION | - |
dc.subject.keywordPlus | INTELLIGIBILITY | - |
dc.subject.keywordPlus | SIMULATION | - |
dc.subject.keywordPlus | SPEECH | - |
dc.subject.keywordAuthor | auscultatory method | - |
dc.subject.keywordAuthor | blood pressure | - |
dc.subject.keywordAuthor | hearing level | - |
dc.subject.keywordAuthor | hearing loss simulator | - |
dc.identifier.url | https://journals.lww.com/bpmonitoring/Fulltext/2014/02000/Does_the_accuracy_of_blood_pressure_measurement.3.aspx | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
222, Wangsimni-ro, Seongdong-gu, Seoul, 04763, Korea+82-2-2220-1365
COPYRIGHT © 2021 HANYANG UNIVERSITY.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.