Bilateral condyle fracture of tibial insert in mobile bearing total knee arthroplasty

Authors
Yoon, Jung-RoJeong, Hyeon-IlOh, Kwang-JunYang, Jae Hyuk
Issue Date
Jan-2014
Publisher
ELSEVIER SCIENCE BV
Keywords
Mobile bearing; Polyethylene breakage; Total knee arthroplasty
Citation
KNEE, v.21, no.1, pp.318 - 321
Indexed
SCIE
SCOPUS
Journal Title
KNEE
Volume
21
Number
1
Start Page
318
End Page
321
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/160818
DOI
10.1016/j.knee.2012.12.014
ISSN
0968-0160
Abstract
We report a case of polyethylene insert breakage in a 45-year-old man after 3.5 years of cruciate retaining type mobile bearing total knee arthroplasty (TKA). Interstingly, both condyles of the polyethylene insert have fractured. The visual assessment done by stereoscopic microscope in the investigation report suggested that the fracture propagation was a result of cyclic loading and that the fracture was from the articular surface as a result of tibio-femoral and anteroposterior shear loading. The initial flexion-extension gap mismatch and/or specific Asian habits like kneeling or deep knee bending could have been the possible factors for over-stress for the insert causing this complication. After replacement of the broken insert and modification for daily activity preventing deep knee flexion, the patient obtained complete relief of previous symptoms. It remains unclear whether insert breakage was secondary to polyethylene insufficient design or to the polyethylene material fracture propagation.
Files in This Item
There are no files associated with this item.
Appears in
Collections
서울 의과대학 > 서울 정형외과학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yang, Jae Hyuk photo

Yang, Jae Hyuk
서울 의과대학 (서울 정형외과학교실)
Read more

Altmetrics

Total Views & Downloads

BROWSE