잘못 장착된 기화기로 인한 전신마취 중 각성
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 김경헌 | - |
dc.contributor.author | 조상윤 | - |
dc.contributor.author | 김규남 | - |
dc.contributor.author | 심재황 | - |
dc.contributor.author | 전우재 | - |
dc.contributor.author | 염종훈 | - |
dc.contributor.author | 신우종 | - |
dc.date.accessioned | 2022-10-07T09:33:36Z | - |
dc.date.available | 2022-10-07T09:33:36Z | - |
dc.date.created | 2022-09-19 | - |
dc.date.issued | 2009-06 | - |
dc.identifier.issn | 2005-6419 | - |
dc.identifier.uri | https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/171645 | - |
dc.description.abstract | Awareness with recall after general anesthesia is a relatively infrequent occurrence and this awareness is often associated with significant adverse psychological sequelae, including posttraumatic stress disorder. The occurrence of awareness is often the consequence of the light anesthetic techniques with using minor anesthetic agents. Because the locking lever of the Penlon Sigma Delta Anesthetic vaporizer (sevoflurane) was at the unlock position, disconnection happened between the anesthetic machine and the vaporizer (the SelectatecⓇ vaporizing system). In our two cases, the inhalation agents were not supplied and awareness with recall was identified after surgery. We recommend to check the concentration of the inhalation agents by monitoring the concentration of using endtidal inhalation agents and to measure the hypnotic level of anesthesia via the bispectral index (BIS) and the auditory evoked potential (AEP). | - |
dc.language | 한국어 | - |
dc.language.iso | ko | - |
dc.publisher | 대한마취통증의학회 | - |
dc.title | 잘못 장착된 기화기로 인한 전신마취 중 각성 | - |
dc.title.alternative | Awareness during general anesthesia due to the unlock position of the vaporizer | - |
dc.type | Article | - |
dc.contributor.affiliatedAuthor | 조상윤 | - |
dc.contributor.affiliatedAuthor | 김규남 | - |
dc.contributor.affiliatedAuthor | 전우재 | - |
dc.contributor.affiliatedAuthor | 염종훈 | - |
dc.contributor.affiliatedAuthor | 신우종 | - |
dc.identifier.doi | 10.4097/kjae.2009.56.6.693 | - |
dc.identifier.bibliographicCitation | Korean Journal of Anesthesiology, v.56, no.6, pp.693 - 697 | - |
dc.relation.isPartOf | Korean Journal of Anesthesiology | - |
dc.citation.title | Korean Journal of Anesthesiology | - |
dc.citation.volume | 56 | - |
dc.citation.number | 6 | - |
dc.citation.startPage | 693 | - |
dc.citation.endPage | 697 | - |
dc.type.rims | ART | - |
dc.identifier.kciid | ART001350330 | - |
dc.description.journalClass | 2 | - |
dc.description.isOpenAccess | Y | - |
dc.description.journalRegisteredClass | kci | - |
dc.subject.keywordAuthor | Awareness | - |
dc.subject.keywordAuthor | Disconnection | - |
dc.subject.keywordAuthor | Vaporizer | - |
dc.subject.keywordAuthor | Awareness | - |
dc.subject.keywordAuthor | Disconnection | - |
dc.subject.keywordAuthor | Vaporizer | - |
dc.identifier.url | https://scienceon.kisti.re.kr/srch/selectPORSrchArticle.do?cn=NART50331964&SITE=CLICK | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
222, Wangsimni-ro, Seongdong-gu, Seoul, 04763, Korea+82-2-2220-1365
COPYRIGHT © 2021 HANYANG UNIVERSITY.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.