Novel method of automatic auscultation for blood pressure measurement using pulses in cuff pressure and korotkoff sound

Authors
Park, DKOh, HSKang, JHKim, In YoungChee, YJLee, JS
Issue Date
Sep-2008
Publisher
IEEE
Citation
Computers in Cardiology, v.35, pp.181 - 184
Indexed
SCOPUS
Journal Title
Computers in Cardiology
Volume
35
Start Page
181
End Page
184
URI
https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/171847
DOI
10.1109/CIC.2008.4749007
ISSN
0276-6574
Abstract
In the NIBP (Non-Invasive Blood Pressure) measurement, we need well trained observer and there are differences in reading values between observers. We suggest new automated auscultation algorithms to select the proper epoch of K-sound for each heart beat. The sound for each epoch is converted into frequency domain and total power was calculated. We decided SBP(Systolic Blood Pressure) point when there was abrupt jump from the baseline. DBP (Diastolic blood pressure) was decided when there was the biggest drop in total power of successive beats. For the validation of our suggested method, human observer assessed SBP/DBP according to EHS (European Hypertension Society) recommendation. The mean differences between the automatic method and traditional auscultation method were 2.0 mmHg for SBP and 3.1 mmHg for DBP. We also took the paired t-test with the two results, and there was no statistically significant difference between these two methods.
Files in This Item
Go to Link
Appears in
Collections
서울 의과대학 > 서울 의공학교실 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, In Young photo

Kim, In Young
COLLEGE OF MEDICINE (DEPARTMENT OF BIOMEDICAL ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE