Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Multitracing Capability of Double-Scattering Compton Imager With NaI(Tl) Scintillator Absorber

Full metadata record
DC Field Value Language
dc.contributor.authorSeo, Hee-
dc.contributor.authorKim, Chan Hyeong-
dc.contributor.authorPark, Jin Hyung-
dc.contributor.authorKim, Jong Kyung-
dc.contributor.authorLee, Ju Hahn-
dc.contributor.authorLee, Chun Sik-
dc.contributor.authorKim, Soo Mee-
dc.contributor.authorLee, Jae Sung-
dc.date.accessioned2022-12-20T17:30:09Z-
dc.date.available2022-12-20T17:30:09Z-
dc.date.created2022-08-27-
dc.date.issued2010-06-
dc.identifier.issn0018-9499-
dc.identifier.urihttps://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/174885-
dc.description.abstractThe Compton camera can provide 3-D images of radioactive material distribution based on a single measurement at a fixed position. The Compton camera also can image several different kinds of radioactive materials simultaneously, by means of the "multitracing" capability. In the present study, this multitracing capability was tested for a double-scattering-type Compton camera, or Double-Scattering Compton Imager (DOCI), which utilizes two double-sided silicon strip detectors (DSSDs) and one NaI(Tl) scintillation detector. Our experimental result shows that the Cs-137 and Co-60 gamma sources can be clearly distinguished in 2-D and 3-D Compton images, and that there is no significant interference between the two gamma sources. The imaging resolutions were determined to be 6.2 and 4.7 mm FWHM for the Cs-137 (662 keV) and Co-60 (1332 keV) point sources at 4 cm, respectively. The angular resolutions, determined from the angular resolution measure (ARM) distributions, were 7.3 degrees and 6.5 degrees for the source energies of 662 and 1332 keV, respectively. The DOCI remains under development; its imaging resolution will be further improved with the incorporation of more sophisticated detectors and the related electronics, including a faster scintillation detector (LYSO) and higher-spatial-resolution position-sensitive detectors.-
dc.language영어-
dc.language.isoen-
dc.publisherIEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC-
dc.titleMultitracing Capability of Double-Scattering Compton Imager With NaI(Tl) Scintillator Absorber-
dc.typeArticle-
dc.contributor.affiliatedAuthorKim, Chan Hyeong-
dc.identifier.doi10.1109/TNS.2009.2035806-
dc.identifier.scopusid2-s2.0-77953921777-
dc.identifier.wosid000278812200053-
dc.identifier.bibliographicCitationIEEE TRANSACTIONS ON NUCLEAR SCIENCE, v.57, no.3, pp.1420 - 1425-
dc.relation.isPartOfIEEE TRANSACTIONS ON NUCLEAR SCIENCE-
dc.citation.titleIEEE TRANSACTIONS ON NUCLEAR SCIENCE-
dc.citation.volume57-
dc.citation.number3-
dc.citation.startPage1420-
dc.citation.endPage1425-
dc.type.rimsART-
dc.type.docTypeArticle; Proceedings Paper-
dc.description.journalClass1-
dc.description.isOpenAccessN-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaEngineering-
dc.relation.journalResearchAreaNuclear Science & Technology-
dc.relation.journalWebOfScienceCategoryEngineering, Electrical & Electronic-
dc.relation.journalWebOfScienceCategoryNuclear Science & Technology-
dc.subject.keywordPlusCAMERA-
dc.subject.keywordPlusRECONSTRUCTION-
dc.subject.keywordPlusRESOLUTION-
dc.subject.keywordAuthorCompton camera-
dc.subject.keywordAuthordouble-scattering-
dc.subject.keywordAuthormultitracing-
dc.subject.keywordAuthorNaI(Tl) scintillator-
dc.identifier.urlhttps://ieeexplore.ieee.org/document/5485180-
Files in This Item
Go to Link
Appears in
Collections
서울 공과대학 > 서울 원자력공학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Chan Hyeong photo

Kim, Chan Hyeong
COLLEGE OF ENGINEERING (DEPARTMENT OF NUCLEAR ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE