Cited 0 time in
Performance evaluation of a table-top Compton camera for various detector parameters
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | Lee, Se Hyung | - |
| dc.contributor.author | Seo, Hee | - |
| dc.contributor.author | An, So Hyun | - |
| dc.contributor.author | Lee, Jae Sung | - |
| dc.contributor.author | Kim, Chan Hyeong | - |
| dc.date.accessioned | 2022-12-21T02:54:03Z | - |
| dc.date.available | 2022-12-21T02:54:03Z | - |
| dc.date.issued | 2008-06 | - |
| dc.identifier.issn | 0168-9002 | - |
| dc.identifier.issn | 1872-9576 | - |
| dc.identifier.uri | https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/178584 | - |
| dc.description.abstract | A prototype table-top Compton camera, composed of two small plane-type position-sensitive semiconductor detectors, is under development. The objective of this study was to find the optimal imaging conditions for the Compton camera in order to maximize its performance in terms of imaging resolution and imaging sensitivity. To that end, the performance of the Compton camera was simulated varying several detector parameters (i.e., the photon energy of the source, the geometrical configuration of the component detectors, and the interaction position resolution of the absorber detector), using the GEANT4 detector simulation toolkit. The Compton camera was found to show its highest performance for the photon energy of 364.5 keV (I-131), but also showed a relatively high performance for 511 keV (F-18) and 662 keV (Cs-137). The Compton camera showed its highest performance also when the scatterer and absorber detectors were positioned in parallel, separated by 10cm. Finally, it was found significantly beneficial to the imaging resolution to increase the interaction position resolution of the absorber detector to 0.3 x 0.3 cm(2) in the planar direction and to 0.5cm in the axial direction. | - |
| dc.format.extent | 4 | - |
| dc.language | 영어 | - |
| dc.language.iso | ENG | - |
| dc.publisher | Elsevier BV | - |
| dc.title | Performance evaluation of a table-top Compton camera for various detector parameters | - |
| dc.type | Article | - |
| dc.publisher.location | 네델란드 | - |
| dc.identifier.doi | 10.1016/j.nima.2008.03.029 | - |
| dc.identifier.scopusid | 2-s2.0-44649096255 | - |
| dc.identifier.wosid | 000257529300021 | - |
| dc.identifier.bibliographicCitation | Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, v.591, no.1, pp 88 - 91 | - |
| dc.citation.title | Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | - |
| dc.citation.volume | 591 | - |
| dc.citation.number | 1 | - |
| dc.citation.startPage | 88 | - |
| dc.citation.endPage | 91 | - |
| dc.type.docType | Article; Proceedings Paper | - |
| dc.description.isOpenAccess | N | - |
| dc.description.journalRegisteredClass | scie | - |
| dc.description.journalRegisteredClass | scopus | - |
| dc.relation.journalResearchArea | Instruments & Instrumentation | - |
| dc.relation.journalResearchArea | Nuclear Science & Technology | - |
| dc.relation.journalResearchArea | Physics | - |
| dc.relation.journalWebOfScienceCategory | Instruments & Instrumentation | - |
| dc.relation.journalWebOfScienceCategory | Nuclear Science & Technology | - |
| dc.relation.journalWebOfScienceCategory | Physics, Nuclear | - |
| dc.relation.journalWebOfScienceCategory | Physics, Particles & Fields | - |
| dc.subject.keywordPlus | PHOTON | - |
| dc.subject.keywordPlus | Cameras | - |
| dc.subject.keywordPlus | Monte Carlo methods | - |
| dc.subject.keywordPlus | Photons | - |
| dc.subject.keywordPlus | Semiconductor detectors | - |
| dc.subject.keywordAuthor | Compton camera | - |
| dc.subject.keywordAuthor | medical imaging | - |
| dc.subject.keywordAuthor | position-sensitive detector | - |
| dc.subject.keywordAuthor | Monte Carlo | - |
| dc.subject.keywordAuthor | GEANT4 | - |
| dc.identifier.url | https://www.sciencedirect.com/science/article/pii/S0168900208004087?via%3Dihub | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
222, Wangsimni-ro, Seongdong-gu, Seoul, 04763, Korea+82-2-2220-1366
COPYRIGHT © 2024 HANYANG UNIVERSITY.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.