Prompt gamma measurements for locating the dose falloff region in the proton therapy
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | Min, Chul-Hee | - |
| dc.contributor.author | Kim, Chan Hyeong | - |
| dc.contributor.author | Youn, Min-Young | - |
| dc.contributor.author | Kim, Jong-Won | - |
| dc.date.accessioned | 2022-12-21T10:10:57Z | - |
| dc.date.available | 2022-12-21T10:10:57Z | - |
| dc.date.issued | 2006-10 | - |
| dc.identifier.issn | 0003-6951 | - |
| dc.identifier.issn | 1077-3118 | - |
| dc.identifier.uri | https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/180919 | - |
| dc.description.abstract | The location of the distal falloff in the proton therapy is an important but often uncertain parameter as different tissue elements are traversed by the beam. A multilayered collimator system has been constructed as a practical means to locate the dose ends by measuring prompt gammas. The collimator is designed to moderate and capture fast neutrons and to prevent unwanted gammas from reaching the scintillation detector. The system has been studied using Monte Carlo technique and has been tested in the beam energy range of 100-200 MeV. Measurements clearly indicated correlations between the gamma distributions and the distal falloff regions. | - |
| dc.language | 영어 | - |
| dc.language.iso | ENG | - |
| dc.publisher | American Institute of Physics | - |
| dc.title | Prompt gamma measurements for locating the dose falloff region in the proton therapy | - |
| dc.type | Article | - |
| dc.publisher.location | 미국 | - |
| dc.identifier.doi | 10.1063/1.2378561 | - |
| dc.identifier.scopusid | 2-s2.0-33750744100 | - |
| dc.identifier.wosid | 000241757500134 | - |
| dc.identifier.bibliographicCitation | Applied Physics Letters, v.89, no.18, pp 183517(1) - 183517(3) | - |
| dc.citation.title | Applied Physics Letters | - |
| dc.citation.volume | 89 | - |
| dc.citation.number | 18 | - |
| dc.citation.startPage | 183517(1) | - |
| dc.citation.endPage | 183517(3) | - |
| dc.type.docType | Article | - |
| dc.description.isOpenAccess | N | - |
| dc.description.journalRegisteredClass | scie | - |
| dc.description.journalRegisteredClass | scopus | - |
| dc.relation.journalResearchArea | Physics | - |
| dc.relation.journalWebOfScienceCategory | Physics, Applied | - |
| dc.subject.keywordPlus | CANCER | - |
| dc.subject.keywordPlus | PET | - |
| dc.subject.keywordPlus | RADIATION | - |
| dc.subject.keywordPlus | BEAMS | - |
| dc.identifier.url | https://aip.scitation.org/doi/10.1063/1.2378561 | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
222, Wangsimni-ro, Seongdong-gu, Seoul, 04763, Korea+82-2-2220-1366
COPYRIGHT © 2024 HANYANG UNIVERSITY.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.