Highly Integrated Guidewire Ultrasound Imaging System-on-a-Chip

Full metadata record
DC Field Value Language
dc.contributor.authorLim, Jaemyung-
dc.contributor.authorTekes, Coskun-
dc.contributor.authorArkan, Evren F.-
dc.contributor.authorRezvanitabar, Ahmad-
dc.contributor.authorDegertekin F. Levent-
dc.contributor.authorGhovanloo, Maysam-
dc.date.accessioned2023-08-16T08:12:12Z-
dc.date.available2023-08-16T08:12:12Z-
dc.date.created2023-07-19-
dc.date.issued2020-05-
dc.identifier.issn0018-9200-
dc.identifier.urihttps://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/189312-
dc.description.abstractIn this article, we present a highly integrated guidewire ultrasound (US) imaging system-on-a-chip (GUISoC) for vascular imaging. The SoC consists of a 16-channel US transmitter (Tx) and receiver (Rx) electronics, on-chip power management IC (PMIC), and quadrature sampler. Using a synthetic aperture imaging algorithm, a Tx/Rx pair, connected to capacitive micromachined ultrasound transducers (CMUTs), can be activated at any time. The Tx generates acoustic waves by driving the CMUT, while the Rx picks up the echo signal and amplify it to be delivered through an interconnect that is driven by a buffer. On-chip logic controls the pulsers that generate the high-voltage (HV)-pulse for Tx. An on-chip PMIC provides 1.8-, 5-, 39-, and 44-V supplies and a clock signal from the two interconnects besides GND. A quadrature sampler down-converts the Rx echo signal to baseband, reducing its bandwidth requirement for the output interconnect. The system design, including transimpedance amplifier (TIA) optimization, based on the equivalent circuit of a specific CMUT is presented. The SoC was fabricated by a 0.18- $\mu \text{m}$ HV CMOS process, occupying 1.5-mm(2) active area and consuming 25.2 and 44 mW from 1.8 to 44 V supplies, respectively. The US Tx and Rx show bandwidths of 32-42 and 32.7-37.5 MHz, respectively. The input-referred noise of the system was measured as 9.66 nA in band with 2-m-long 52 American Wire Gauge (AWG) wire interconnects. The functionality of the GUISoC was verified in vitro by imaging wire targets.-
dc.language영어-
dc.language.isoen-
dc.publisherIEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC-
dc.titleHighly Integrated Guidewire Ultrasound Imaging System-on-a-Chip-
dc.typeArticle-
dc.contributor.affiliatedAuthorLim, Jaemyung-
dc.identifier.doi10.1109/JSSC.2020.2967551-
dc.identifier.scopusid2-s2.0-85083990620-
dc.identifier.wosid000530283400016-
dc.identifier.bibliographicCitationIEEE JOURNAL OF SOLID-STATE CIRCUITS, v.55, no.5, pp.1310 - 1323-
dc.relation.isPartOfIEEE JOURNAL OF SOLID-STATE CIRCUITS-
dc.citation.titleIEEE JOURNAL OF SOLID-STATE CIRCUITS-
dc.citation.volume55-
dc.citation.number5-
dc.citation.startPage1310-
dc.citation.endPage1323-
dc.type.rimsART-
dc.type.docType정기학술지(Article(Perspective Article포함))-
dc.description.journalClass1-
dc.description.isOpenAccessY-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaEngineering-
dc.relation.journalWebOfScienceCategoryEngineering, Electrical & Electronic-
dc.subject.keywordPlusCMUTTRANSDUCERSDESIGNARRAYSIVUS-
dc.subject.keywordAuthorCapacitive micromachined ultrasound transducer (CMUT)intravascular ultrasound (IVUS)quadrature samplingultrasound (US) imaging-
dc.identifier.urlhttps://ieeexplore.ieee.org/document/8976311-
Files in This Item
Go to Link
Appears in
Collections
서울 공과대학 > 서울 융합전자공학부 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lim, Jaemyung photo

Lim, Jaemyung
COLLEGE OF ENGINEERING (SCHOOL OF ELECTRONIC ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE