Cited 0 time in
Detemination of Distal Dose Edge with Array-type Prompt Gamma Detection System for 39 MeV Proton Beam
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | 김찬형 | - |
| dc.date.accessioned | 2021-08-03T23:21:46Z | - |
| dc.date.available | 2021-08-03T23:21:46Z | - |
| dc.date.issued | 2008-09-04 | - |
| dc.identifier.uri | https://scholarworks.bwise.kr/hanyang/handle/2021.sw.hanyang/63984 | - |
| dc.description.abstract | To verify experimentally the distal dose edge or proton beam range, an array-type prompt gamma measurement system was constructed and applied to a 39 MeV proton beam. The proton beam was delivered to a water phantom and then the prompt gammas, generated in the passage of the proton beam were measured with an array-type prompt gamma detection system` The measurement system Is composed of a collimator with multiple collimation slits, an CsI(Tl) scintillator and a photodiode, and lead shielding` The distal dose edge was also measured with a plane-parallel ionization chamber and, then, the result was compared with the distribution of the prompt gammas. The prompt gamma energy spectra around the distal dose edge was measured and compared for the validity of the energy window of 4-10 MeV. The distribution of the prompt gammas, measured by the array-type prompt gamma detection system, sharply decreased at the distal dose edge, within the range of a few millimeters. The effect of the background gammas was not very significant at least [01` the 39 MeV proton beam case. The results were very encouraging, and demonstrated that the distal dose of the proton beam in the patient can be determined by directly measuring the distribution of the prompt gammas with an array-type prompt gamma detection system. | - |
| dc.title | Detemination of Distal Dose Edge with Array-type Prompt Gamma Detection System for 39 MeV Proton Beam | - |
| dc.type | Conference | - |
| dc.citation.conferenceName | The 12th International Workshop on Accelerator and Beam Utillization | - |
| dc.citation.conferencePlace | Hyundai Hotel,Gyeongju Korea | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
222, Wangsimni-ro, Seongdong-gu, Seoul, 04763, Korea+82-2-2220-1366
COPYRIGHT © 2024 HANYANG UNIVERSITY.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.