Performance evaluation of SimPET-L and SimPET-XL: MRI-compatible small-animal PET systems with rat-body imaging capabilityopen access
- Authors
- Seo, Minjee; Ko, Guen Bae; Kim, Kyeong Yun; Son, Jeong-Whan; Byun, Jung Woo; Lee, Yun-Sang; Kim, Kyeong Min; Park, Jang Woo; Kim, Kipom; Lee, Taekwan; Lee, Jae Sung
- Issue Date
- Mar-2023
- Publisher
- Springer International Publishing AG
- Keywords
- PET; MRI; Molecular imaging; NEMA performance; Instrumentation
- Citation
- EJNMMI Physics, v.10, no.1
- Journal Title
- EJNMMI Physics
- Volume
- 10
- Number
- 1
- URI
- http://scholarworks.bwise.kr/kbri/handle/2023.sw.kbri/144
- DOI
- 10.1186/s40658-023-00534-x
- ISSN
- 2197-7364
- Abstract
- BackgroundSimPET-L and SimPET-XL have recently been introduced with increased transaxial fields of view (FOV) compared with their predecessors (SimPET (TM) and SimPET-X), enabling whole-body positron emission tomography (PET) imaging of rats. We conducted performance evaluations of SimPET-L and SimPET-XL and rat-body imaging with SimPET-XL to demonstrate the benefits of increased axial and transaxial FOVs.ProceduresThe detector blocks in SimPET-L and SimPET-XL consist of two 4 x 4 silicon photomultiplier arrays coupled with 20 x 9 array lutetium oxyorthosilicate crystals. SimPET-L and SimPET-XL have an inner diameter (bore size) of 7.6 cm, and they are composed of 40 and 80 detector blocks yielding axial lengths of 5.5 and 11 cm, respectively. Each system was evaluated according to the National Electrical Manufacturers Association NU4-2008 protocol. Rat imaging studies, such as F-18-NaF and F-18-FDG PET, were performed using SimPET-XL.ResultsThe radial resolutions at the axial center measured using the filtered back projection, 3D ordered-subset expectation maximization (OSEM), and 3D OSEM with point spread functions correction were 1.7, 0.82, and 0.82 mm FWHM in SimPET-L and 1.7, 0.91, and 0.91 mm FWHM in SimPET-XL, respectively. The peak sensitivities of SimPET-L and SimPET-XL were 6.30% and 10.4% for an energy window of 100-900 keV and 4.44% and 7.25% for a window of 250-750 keV, respectively. The peak noise equivalent count rate with an energy window of 250-750 keV was 249 kcps at 44.9 MBq for SimPET-L and 349 kcps at 31.3 MBq for SimPET-XL. In SimPET-L, the uniformity was 4.43%, and the spill-over ratios in air- and water-filled chambers were 5.54% and 4.10%, respectively. In SimPET-XL, the uniformity was 3.89%, and the spill-over ratio in the air- and water-filled chambers were 3.56% and 3.60%. Moreover, SimPET-XL provided high-quality images of rats.ConclusionSimPET-L and SimPET-XL show adequate performance compared with other SimPET systems. In addition, their large transaxial and long axial FOVs provide imaging capability for rats with high image quality.
- Files in This Item
- There are no files associated with this item.
- Appears in
Collections - 연구전략실 > 인프라구축팀 > 1. Journal Articles
- 연구전략실 > 첨단뇌연구장비센터 > 1. Journal Articles
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.