Detailed Information

Cited 0 time in webofscience Cited 149 time in scopus
Metadata Downloads

Accuracy and Precision of Continuous Noninvasive Arterial Pressure Monitoring Compared with Invasive Arterial Pressure A Systematic Review and Meta-analysis

Full metadata record
DC Field Value Language
dc.contributor.authorKim, Sang-Hyun-
dc.contributor.authorLilot, Marc-
dc.contributor.authorSidhu, Kulraj S.-
dc.contributor.authorRinehart, Joseph-
dc.contributor.authorYu, Zhaoxia-
dc.contributor.authorCanales, Cecilia-
dc.contributor.authorCannesson, Maxime-
dc.date.accessioned2021-08-11T22:46:35Z-
dc.date.available2021-08-11T22:46:35Z-
dc.date.issued2014-05-
dc.identifier.issn0003-3022-
dc.identifier.issn1528-1175-
dc.identifier.urihttps://scholarworks.bwise.kr/sch/handle/2021.sw.sch/12234-
dc.description.abstractBackground: Continuous noninvasive arterial pressure monitoring devices are available for bedside use, but the accuracy and precision of these devices have not been evaluated in a systematic review and meta-analysis. Methods: The authors performed a systematic review and meta-analysis of studies comparing continuous noninvasive arterial pressure monitoring with invasive arterial pressure monitoring. Random-effects pooled bias and SD of bias for systolic arterial pressure, diastolic arterial pressure, and mean arterial pressure were calculated. Continuous noninvasive arterial pressure monitoring was considered acceptable if pooled estimates of bias and SD were not greater than 5 and 8 mmHg, respectively, as recommended by the Association for the Advancement of Medical Instrumentation. Results: Twenty-eight studies (919 patients) were included. The overall random-effect pooled bias and SD were -1.6 12.2 mmHg (95% limits of agreement -25.5 to 22.2 mmHg) for systolic arterial pressure, 5.3 +/- 8.3 mmHg (-11.0 to 21.6 mmHg) for diastolic arterial pressure, and 3.2 +/- 8.4 mmHg (-13.4 to 19.7 mmHg) for mean arterial pressure. In 14 studies focusing on currently commercially available devices, bias and SD were -1.8 +/- 12.4 mmHg (-26.2 to 22.5 mmHg) for systolic arterial pressure, 6.0 +/- 8.6 mmHg (-10.9 to 22.9 mmHg) for diastolic arterial pressure, and 3.9 +/- 8.7 mmHg (-13.1 to 21.0 mmHg) for mean arterial pressure. Conclusions: The results from this meta-analysis found that inaccuracy and imprecision of continuous noninvasive arterial pressure monitoring devices are larger than what was defined as acceptable. This may have implications for clinical situations where continuous noninvasive arterial pressure is being used for patient care decisions.-
dc.format.extent18-
dc.language영어-
dc.language.isoENG-
dc.publisherLippincott Williams & Wilkins Ltd.-
dc.titleAccuracy and Precision of Continuous Noninvasive Arterial Pressure Monitoring Compared with Invasive Arterial Pressure A Systematic Review and Meta-analysis-
dc.typeArticle-
dc.publisher.location미국-
dc.identifier.doi10.1097/ALN.0000000000000226-
dc.identifier.scopusid2-s2.0-84899489146-
dc.identifier.wosid000335175100008-
dc.identifier.bibliographicCitationAnesthesiology, v.120, no.5, pp 1080 - 1097-
dc.citation.titleAnesthesiology-
dc.citation.volume120-
dc.citation.number5-
dc.citation.startPage1080-
dc.citation.endPage1097-
dc.type.docTypeArticle-
dc.description.isOpenAccessN-
dc.description.journalRegisteredClasssci-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaAnesthesiology-
dc.relation.journalWebOfScienceCategoryAnesthesiology-
dc.subject.keywordPlusCRITICALLY-ILL PATIENTS-
dc.subject.keywordPlusRISK SURGICAL-PATIENTS-
dc.subject.keywordPlusFINGER BLOOD-PRESSURE-
dc.subject.keywordPlusCLINICAL-EVALUATION-
dc.subject.keywordPlusGENERAL-ANESTHESIA-
dc.subject.keywordPlusCARDIAC-SURGERY-
dc.subject.keywordPlusTONOMETRY-
dc.subject.keywordPlusDEVICE-
dc.subject.keywordPlusAGREEMENT-
dc.subject.keywordPlusFINAPRES-
dc.subject.keywordAuthor마취과학-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Anesthesiology > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Sang hyun photo

Kim, Sang hyun
College of Medicine (Department of Anesthesiology)
Read more

Altmetrics

Total Views & Downloads

BROWSE