Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Efficacy of noninvasive pulse co-oximetry as compared to invasive laboratory-based hemoglobin measurement during spinal anesthesia

Full metadata record
DC Field Value Language
dc.contributor.author정진헌-
dc.contributor.author유시현-
dc.contributor.author지재영-
dc.contributor.author김난설-
dc.contributor.author서용한-
dc.contributor.author공형윤-
dc.contributor.author김재우-
dc.contributor.author김종분-
dc.date.accessioned2021-08-11T23:44:52Z-
dc.date.available2021-08-11T23:44:52Z-
dc.date.issued2014-
dc.identifier.issn1975-5171-
dc.identifier.issn2383-7977-
dc.identifier.urihttps://scholarworks.bwise.kr/sch/handle/2021.sw.sch/12713-
dc.description.abstractBackground: The Masimo Radical 7 (Masimo Corp., Irvine, CA,USA) pulse co-oximeterⓇ noninvasively determines the hemoglobinconcentration using the principle of transcutaneous spectrophotometry. We compared hemoglobin levels determined using thisdevice (SpHb) with those determined using an invasive laboratory-based technique (tHb) during spinal anesthesia. Methods: Thirty patients received spinal anesthesia with 0.5%hyperbaric bupivacaine. The pulse co-oximeter probe was mountedon the second toe, and arterial blood samples were obtained froma radial artery catheter. SpHb, tHb, and perfusion index (PI) valueswere recorded before and 20 and 40 min after intrathecal injectionof bupivacaine. Results: Before spinal anesthesia, the SpHb and tHb showed asignificant difference of −2.86 ± 1.56 g/dl (P < 0.005), but nosignificant differences were found between tHb and SpHb at 20 and40 min after spinal anesthesia (−0.16 ± 2.45 g/dl and 0.29 ± 2.68g/dl). Additionally, PI was significantly increased at 20 and 40 minafter spinal anesthesia compared to the pre-anesthetic value (P <0.001). Conclusions: The toe is not the monitoring site for pulseco-oximetry in adult patients, but the pulse co-oximetry on the toeappears to be appropriate as a noninvasive hemoglobin monitoringdevice after spinal anesthesia.-
dc.format.extent5-
dc.language영어-
dc.language.isoENG-
dc.publisher대한마취통증의학회-
dc.titleEfficacy of noninvasive pulse co-oximetry as compared to invasive laboratory-based hemoglobin measurement during spinal anesthesia-
dc.title.alternativeEfficacy of noninvasive pulse co-oximetry as compared to invasive laboratory-based hemoglobin measurement during spinal anesthesia-
dc.typeArticle-
dc.publisher.location대한민국-
dc.identifier.bibliographicCitationAnesthesia and Pain Medicine, v.9, no.4, pp 277 - 281-
dc.citation.titleAnesthesia and Pain Medicine-
dc.citation.volume9-
dc.citation.number4-
dc.citation.startPage277-
dc.citation.endPage281-
dc.identifier.kciidART001922759-
dc.description.isOpenAccessN-
dc.description.journalRegisteredClasskciCandi-
dc.subject.keywordAuthorHemoglobin-
dc.subject.keywordAuthorPulse co-oximetry-
dc.subject.keywordAuthorSpinal anesthesia-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Anesthesiology > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Seo, Young Han photo

Seo, Young Han
College of Medicine (Department of Anesthesiology)
Read more

Altmetrics

Total Views & Downloads

BROWSE