Impaction of a continuous glucose monitoring sensor

Full metadata record
DC Field Value Language
dc.contributor.authorPark, Kyong Chan-
dc.contributor.authorChoi, Hwan Jun-
dc.date.accessioned2021-09-10T05:44:55Z-
dc.date.available2021-09-10T05:44:55Z-
dc.date.issued2021-07-
dc.identifier.issn2234-6163-
dc.identifier.issn2234-6171-
dc.identifier.urihttps://scholarworks.bwise.kr/sch/handle/2021.sw.sch/18704-
dc.description.abstractA 33-year-old man presented to the plastic surgery department for foreign body removal 1 month after the insertion of a continuous glucose monitoring (CGM) sensor (Dexcom G5) in the left upper arm. The patient had used the CGM system for 5 years, and the insertion was done in the usual manner. The entire sensor wire was visible on simple radiography and ultrasonography. In the operating room, and the sensor wire was identified in the intermuscular septum and removed. No foreign body reaction or inflammatory signs were found around the CGM, and the extracted wire measured 2.5 cm. Thus, it was assumed that the whole sensor wire was detached from the transmitter, not fractured. No remnant foreign body was observed on follow-up simple radiography.-
dc.format.extent3-
dc.language영어-
dc.language.isoENG-
dc.publisher대한성형외과학회-
dc.titleImpaction of a continuous glucose monitoring sensor-
dc.typeArticle-
dc.publisher.location대한민국-
dc.identifier.doi10.5999/aps.2021.00178-
dc.identifier.scopusid2-s2.0-85111744287-
dc.identifier.wosid000674975200010-
dc.identifier.bibliographicCitationArchives of Plastic Surgery, v.48, no.4, pp 392 - 394-
dc.citation.titleArchives of Plastic Surgery-
dc.citation.volume48-
dc.citation.number4-
dc.citation.startPage392-
dc.citation.endPage394-
dc.type.docTypeArticle-
dc.identifier.kciidART002739316-
dc.description.isOpenAccessY-
dc.description.journalRegisteredClassscopus-
dc.description.journalRegisteredClassesci-
dc.description.journalRegisteredClasskci-
dc.relation.journalResearchAreaSurgery-
dc.relation.journalWebOfScienceCategorySurgery-
dc.subject.keywordAuthorBiosensor-
dc.subject.keywordAuthorHome blood glucose monitoring-
dc.subject.keywordAuthorForeign bodies-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Plastic Surgery > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Choi, Hwan jun photo

Choi, Hwan jun
College of Medicine (Department of Plastic Surgery)
Read more

Altmetrics

Total Views & Downloads

BROWSE