Detailed Information

Cited 0 time in webofscience Cited 5 time in scopus
Metadata Downloads

Immunosuppressive Drug Measurement by Liquid Chromatography Coupled to Tandem Mass Spectrometry: Interlaboratory Comparison in the Korean Clinical Laboratories

Full metadata record
DC Field Value Language
dc.contributor.authorKim, Hyun-Ki-
dc.contributor.authorPark, Hyung-Doo-
dc.contributor.authorLee, Sang-Guk-
dc.contributor.authorChae, Hyojin-
dc.contributor.authorSong, Sang Hoon-
dc.contributor.authorLee, Yong-Wha-
dc.contributor.authorYun, Yeo-Min-
dc.contributor.authorAhn, Sunhyun-
dc.contributor.authorKim, Serim-
dc.contributor.authorLee, Sun Min-
dc.contributor.authorLee, Soo-Youn-
dc.contributor.authorChun, Sail-
dc.date.accessioned2021-09-10T05:49:10Z-
dc.date.available2021-09-10T05:49:10Z-
dc.date.issued2021-05-
dc.identifier.issn2234-3806-
dc.identifier.issn2234-3814-
dc.identifier.urihttps://scholarworks.bwise.kr/sch/handle/2021.sw.sch/18891-
dc.description.abstractBackground: Liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) is increasingly used for immunosuppressive drug tests. However, most LC-MS/MS tests are laboratory-developed and their agreement is unknown in different Korean laboratories. This interlaboratory comparison study evaluated test reproducibility and identified potential error sources. Methods: Test samples containing three concentrations of tacrolimus, sirolimus, everolimus, cyclosporine, and mycophenolic acid were prepared by pooling surplus samples from patients undergoing routine therapeutic drug monitoring and tested in duplicate in the participating 10 clinical laboratories. Reconstitution and storage experiments were conducted for the commonly used commercial calibrator set. The robust estimators of reproducibility parameters were calculated. Spearman's rank correlation coefficient (rho, rho) was used to evaluate the correlation between drugs. Multiple linear regression was used to determine whether the experimental conditions alter the calibration curves. Results: The reproducibility coefficient of variation exceeded 10% only for sirolimus concentrations 1 and 2 (10.8% and 12.5%, respectively) and everolimus concentrations 1 and 2 (12.3% and 11.4%, respectively). The percent difference values showed weak correlations between sirolimus and everolimus (rho=0.334, P=0.175). The everolimus calibration curve slope was significantly altered after reconstitution following prolonged 5 degrees C storage (P=0.015 for 14 days; P=0.025 for 28 days); the expected differences at 6 ng/mL were 0.598% for 14 days and 0.384% for 28 days. Conclusions: LC-MS/MS test reproducibility for immunosuppressive drugs seems to be good in the Korean clinical laboratories. Continuous efforts are required to achieve test standardization and harmonization, especially for sirolimus and everolimus.-
dc.format.extent9-
dc.language영어-
dc.language.isoENG-
dc.publisher대한진단검사의학회-
dc.titleImmunosuppressive Drug Measurement by Liquid Chromatography Coupled to Tandem Mass Spectrometry: Interlaboratory Comparison in the Korean Clinical Laboratories-
dc.typeArticle-
dc.publisher.location대한민국-
dc.identifier.doi10.3343/alm.2021.41.3.268-
dc.identifier.scopusid2-s2.0-85098016858-
dc.identifier.wosid000598521800002-
dc.identifier.bibliographicCitationAnnals of Laboratory Medicine, v.41, no.3, pp 268 - 276-
dc.citation.titleAnnals of Laboratory Medicine-
dc.citation.volume41-
dc.citation.number3-
dc.citation.startPage268-
dc.citation.endPage276-
dc.type.docTypeArticle-
dc.description.isOpenAccessN-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.description.journalRegisteredClasskci-
dc.relation.journalResearchAreaMedical Laboratory Technology-
dc.relation.journalWebOfScienceCategoryMedical Laboratory Technology-
dc.subject.keywordPlusLC-MS-
dc.subject.keywordPlusPERFORMANCE-
dc.subject.keywordPlusSTANDARDIZATION-
dc.subject.keywordPlusTACROLIMUS-
dc.subject.keywordAuthorImmunosuppressive drug-
dc.subject.keywordAuthorTherapeutic drug monitoring-
dc.subject.keywordAuthorLiquid chromatography coupled to tandem mass spectrometry-
dc.subject.keywordAuthorInterlaboratory comparison-
dc.subject.keywordAuthorSirolimus-
dc.subject.keywordAuthorEverolimus-
dc.subject.keywordAuthorReproducibility-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Clinical Pathology > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Yong Wha photo

Lee, Yong Wha
College of Medicine (Department of Clinical Pathology)
Read more

Altmetrics

Total Views & Downloads

BROWSE