Unstable Vancouver B1 periprosthetic femoral fracture fixation: A biomechanical comparison between a novel C-shaped memory alloy implant and cerclage wiringopen access

Authors
Oh, Seog-HyunSuh, Yu-SungEghan-Acquah, EmmanuelYurevich, Kollerov MikhailWon, Sung-HunBaek, Min JungLee, Sung-JaeChoi, Sung-Woo
Issue Date
Mar-2024
Publisher
SAGE PUBLICATIONS LTD
Keywords
Femur; Fracture-Periprosthetic; Arthroplasty; Biomechanical Phenomena; Cable; Cerclage wire
Citation
JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, v.52, no.3
Journal Title
JOURNAL OF INTERNATIONAL MEDICAL RESEARCH
Volume
52
Number
3
URI
https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/26424
DOI
10.1177/03000605241240946
ISSN
0300-0605
1473-2300
Abstract
Background To compare the biomechanical stability of a novel, C-shaped nickel-titanium shape memory alloy (SMA) implant (C-clip) with traditional cerclage wiring in the fixation of a Vancouver B1 (VB1) periprosthetic femoral fracture (PFF).Methods In total, 18 synthetic femoral fracture models were constructed to obtain unstable VB1 fracture with an oblique fracture line 8 cm below the lesser trochanter. For each model, the distal portion was repaired using a 10-hole locking plate and four distal bi-cortical screws. The proximal portion was repaired using either three, threaded cerclage wirings or three, novel C-shaped implants. Specimens underwent biomechanical testing using axial compression, torsional and four-point bending tests. Each test was performed on three specimens.Results The C-clip was statistically significantly stronger (i.e., stiffer) than cerclage wiring in the three biomechanical tests. For axial compression, medians (ranges) were 39 (39-41) and 35 (35-35) N/mm, for the C-clip and cerclage wiring, respectively. For torsion, medians (ranges) were, 0.44 (0.44-0.45) and 0.30 (0.30-0.33) N/mm for the C-clip and cerclage wiring, respectively. For the four-point bending test, medians (ranges) were 39 (39-41) and 28 (28-31) N/mm; for the C-clip and cerclage wiring, respectively.Conclusion Results from this small study show that the novel, C-shaped SMA appears to be biomechanically superior to traditional cerclage wiring in terms of stiffness, axial compression, torsion and four-point bending, and may be a valuable alternative in the repair of VB1 PFF. Further research is necessary to support these results.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Orthopedic Surgery > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Choi, Sung Woo photo

Choi, Sung Woo
College of Medicine (Department of Orthopedic Surgery)
Read more

Altmetrics

Total Views & Downloads

BROWSE