Measurement reliability of automated oscillometric blood pressure monitor in the elderly with atrial fibrillation: a systematic review and meta-analysis
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Park, Seong-Hi | - |
dc.contributor.author | Choi, Yun-Kyoung | - |
dc.date.accessioned | 2021-08-11T08:37:48Z | - |
dc.date.available | 2021-08-11T08:37:48Z | - |
dc.date.issued | 2020-02 | - |
dc.identifier.issn | 1359-5237 | - |
dc.identifier.issn | 1473-5725 | - |
dc.identifier.uri | https://scholarworks.bwise.kr/sch/handle/2021.sw.sch/3124 | - |
dc.description.abstract | Objectives This study aimed to identify whether automated oscillometric blood pressure monitor (AOBPM) is a reliable blood pressure (BP) measurement tool in geriatric patients with atrial fibrillation (AF) with high variability in BP and to evaluate whether it can be applied in practice. Methods Electronic searches were performed in databases including MEDLINE, EMBASE, the Cochrane Library, and CINAHL by using the following keywords: 'atrial fibrillation,' 'atrial flutter, 'blood pressure monitor', 'sphygmomanometer.' The QUADAS-2 was applied to assess the internal validity of selected studies. Meta-analysis was performed using RevMan 5.3 program. Design: Systematic review. Results We identified 10 studies, including 938 geriatric patients with AF. We compared with the previously used BP measurement method (mainly office) and AOBPM, and the patients with AF were divided into the AF-AF (atrial fibrillation rhythm continued) and AF-SR groups (sinus rhythm recovered). The difference in the systolic BP was -3.0 mmHg [95% confidence interval (CI): -6.58 to 0.59] and -1.62 (95% CI: -6.08 to 2.84) mmHg in the AF-AF and AF-SR groups, respectively. The difference in the diastolic BP was 0.17 (95% CI: -2.90 to 3.25) mmHg and -0.23 (95% CI: -5.11 to 4.65) mmHg, respectively. Conclusion This review showed that the BP difference from AOBPM compared with the auscultatory BP method was less than 5 mmHg in the elderly with AF. This difference is acceptable in clinical practice. However, AOBPM compared with invasive arterial BP in the diastolic BP was a difference of 5 mmHg or more, and so its accuracy cannot be assured. | - |
dc.format.extent | 11 | - |
dc.language | 영어 | - |
dc.language.iso | ENG | - |
dc.publisher | Lippincott Williams & Wilkins Ltd. | - |
dc.title | Measurement reliability of automated oscillometric blood pressure monitor in the elderly with atrial fibrillation: a systematic review and meta-analysis | - |
dc.type | Article | - |
dc.publisher.location | 미국 | - |
dc.identifier.doi | 10.1097/MBP.0000000000000414 | - |
dc.identifier.scopusid | 2-s2.0-85077224009 | - |
dc.identifier.wosid | 000528021400002 | - |
dc.identifier.bibliographicCitation | Blood Pressure Monitoring, v.25, no.1, pp 2 - 12 | - |
dc.citation.title | Blood Pressure Monitoring | - |
dc.citation.volume | 25 | - |
dc.citation.number | 1 | - |
dc.citation.startPage | 2 | - |
dc.citation.endPage | 12 | - |
dc.type.docType | Review | - |
dc.description.isOpenAccess | N | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
dc.relation.journalResearchArea | Cardiovascular System & Cardiology | - |
dc.relation.journalWebOfScienceCategory | Peripheral Vascular Disease | - |
dc.subject.keywordPlus | MEASURING DEVICES | - |
dc.subject.keywordPlus | ANEROID SPHYGMOMANOMETERS | - |
dc.subject.keywordPlus | HEART-ASSOCIATION | - |
dc.subject.keywordPlus | HYPERTENSION | - |
dc.subject.keywordPlus | MANAGEMENT | - |
dc.subject.keywordPlus | SOCIETY | - |
dc.subject.keywordPlus | HOME | - |
dc.subject.keywordPlus | RECOMMENDATIONS | - |
dc.subject.keywordPlus | VALIDATION | - |
dc.subject.keywordPlus | DIAGNOSIS | - |
dc.subject.keywordAuthor | aged | - |
dc.subject.keywordAuthor | atrial fibrillation | - |
dc.subject.keywordAuthor | blood pressure determination | - |
dc.subject.keywordAuthor | sphygmomanometer | - |
dc.subject.keywordAuthor | systematic review | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
(31538) 22, Soonchunhyang-ro, Asan-si, Chungcheongnam-do, Republic of Korea+82-41-530-1114
COPYRIGHT 2021 by SOONCHUNHYANG UNIVERSITY ALL RIGHTS RESERVED.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.