Mid-term results of the BIOLOX delta ceramic-on-ceramic total hip arthroplasty

Full metadata record
DC Field Value Language
dc.contributor.authorLee, Y. K.-
dc.contributor.authorHa, Y. C.-
dc.contributor.authorYoo, J-I.-
dc.contributor.authorJo, W. L.-
dc.contributor.authorKim, K-C.-
dc.contributor.authorKoo, K. H.-
dc.date.available2019-03-08T08:38:18Z-
dc.date.issued2017-06-
dc.identifier.issn2049-4394-
dc.identifier.issn2049-4408-
dc.identifier.urihttps://scholarworks.bwise.kr/cau/handle/2019.sw.cau/4359-
dc.description.abstractAims We conducted a prospective study of a delta ceramic total hip arthroplasty (THA) to determine the rate of ceramic fracture, to characterise post-operative noise, and to evaluate the mid-term results and survivorship. Patients and Methods Between March 2009 and March 2011, 274 patients (310 hips) underwent cementless THA using a delta ceramic femoral head and liner. At each follow-up, clinical and radiological outcomes were recorded. A Kaplan-Meier analysis was undertaken to estimate survival. Results Four patients (four hips) died and 18 patients (20 hips) were lost to follow-up within five years. The remaining 252 patients (286 hips) were followed for a mean of 66.5 months (60 to 84). There were 144 men (166 hips) and 108 women (120 hips) with a mean age of 49.7 years (16 to 83) at surgery. The mean pre-operative Harris Hip Score of 47.1 points improved to 93.8 points at final follow-up. Six patients reported squeaking in seven hips; however, none were audible. Radiolucent lines involving Gruen zones one and/or seven were seen in 52 hips (18.2%). No hip had detectable wear, focal osteolysis or signs of loosening. One hip was revised because of fracture of the ceramic liner, which occurred due to an undetected malseating of the ceramic liner at the time of surgery. One hip was revised for a periprosthetic fracture of the femur, and one hip was treated for periprosthetic joint infection. The six-year survivorship with re-operation for any reason as the endpoint was 99.0% (95% confidence interval 97.8% to 100%). Discussion The rate of delta ceramic fracture was 0.3% (one of 286). While ceramic head fracture was dominant in previous ceramic-on-ceramic THA, fracture of the delta ceramic liner due to malseating is a concern.-
dc.format.extent8-
dc.language영어-
dc.language.isoENG-
dc.publisherBRITISH EDITORIAL SOC BONE JOINT SURGERY-
dc.titleMid-term results of the BIOLOX delta ceramic-on-ceramic total hip arthroplasty-
dc.typeArticle-
dc.identifier.doi10.1302/0301-620X.99B6.BJJ-2016-0486-
dc.identifier.bibliographicCitationBONE & JOINT JOURNAL, v.99B, no.6, pp 741 - 748-
dc.description.isOpenAccessN-
dc.identifier.wosid000402402100006-
dc.identifier.scopusid2-s2.0-85021256689-
dc.citation.endPage748-
dc.citation.number6-
dc.citation.startPage741-
dc.citation.titleBONE & JOINT JOURNAL-
dc.citation.volume99B-
dc.type.docTypeArticle-
dc.publisher.location영국-
dc.subject.keywordPlusFEMORAL-HEAD-
dc.subject.keywordPlusACETABULAR COMPONENTS-
dc.subject.keywordPlusFOLLOW-UP-
dc.subject.keywordPlusALUMINA-
dc.subject.keywordPlusREPLACEMENT-
dc.subject.keywordPlusBEARINGS-
dc.subject.keywordPlusFRACTURE-
dc.subject.keywordPlusMINIMUM-
dc.subject.keywordPlusSURFACE-
dc.subject.keywordPlusOSTEONECROSIS-
dc.relation.journalResearchAreaOrthopedics-
dc.relation.journalResearchAreaSurgery-
dc.relation.journalWebOfScienceCategoryOrthopedics-
dc.relation.journalWebOfScienceCategorySurgery-
dc.description.journalRegisteredClasssci-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > College of Medicine > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE