Smartwatches as chest compression feedback devices: A feasibility study
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Song, Yeongtak | - |
dc.contributor.author | Chee, Youngjoon | - |
dc.contributor.author | Oh, Jaehoon | - |
dc.contributor.author | Ahn, Chiwon | - |
dc.contributor.author | Lim, Tae Ho | - |
dc.date.accessioned | 2024-01-18T05:01:27Z | - |
dc.date.available | 2024-01-18T05:01:27Z | - |
dc.date.issued | 2016-06 | - |
dc.identifier.issn | 0300-9572 | - |
dc.identifier.issn | 1873-1570 | - |
dc.identifier.uri | https://scholarworks.bwise.kr/cau/handle/2019.sw.cau/71054 | - |
dc.description.abstract | Background: Recently, there have been attempts to use smartphones and smartwatches as the feedback devices to improve the quality of chest compressions. In this study, we compared chest compression depth feedback accuracy between a smartphone and a smartwatch in a hands-only cardiopulmonary resuscitation scenario, using a manikin with a displacement sensor system. Methods: Ten basic life support providers participated in this study. Guided by the chest compression depths displayed on the monitor of a laptop, which received data from the manikin, each participant performed 2 min of chest compressions for each target depth (35 mm and 55 mm) on a manikin while gripping a smartphone and wearing a smartwatch. Participants had a rest of 1 h between the instances, and the first target depth was set at random. Each chest compression depth data value from the smartphone and smartwatch and a corresponding reference value from the manikin with the displacement system were recorded. To compare the accuracy between the smartphone and smartwatch, the errors, expressed as the absolute of the differences between the reference and each device, were calculated. Results: At both target depths, the error of the smartwatch were significantly smaller than that of the smartphone (the errors of the smartphone vs. smartwatch at 35 mm: 3.4 (1.3) vs. 2.1 (0.8) mm; p = 0.008; at 55 mm: 5.3 (2.8) vs. 2.3 (0.9) mm; p = 0.023). Conclusion: The smartwatch-based chest compression depth feedback was more accurate than smartphone-based feedback. (C) 2016 Elsevier Ireland Ltd. All rights reserved. | - |
dc.format.extent | 4 | - |
dc.language | 영어 | - |
dc.language.iso | ENG | - |
dc.publisher | ELSEVIER IRELAND LTD | - |
dc.title | Smartwatches as chest compression feedback devices: A feasibility study | - |
dc.type | Article | - |
dc.identifier.doi | 10.1016/j.resuscitation.2016.03.014 | - |
dc.identifier.bibliographicCitation | RESUSCITATION, v.103, pp 20 - 23 | - |
dc.description.isOpenAccess | N | - |
dc.identifier.wosid | 000375898600016 | - |
dc.identifier.scopusid | 2-s2.0-84961924284 | - |
dc.citation.endPage | 23 | - |
dc.citation.startPage | 20 | - |
dc.citation.title | RESUSCITATION | - |
dc.citation.volume | 103 | - |
dc.type.docType | Article | - |
dc.publisher.location | 아일랜드 | - |
dc.subject.keywordAuthor | CPR | - |
dc.subject.keywordAuthor | Chest compression | - |
dc.subject.keywordAuthor | Feedback device | - |
dc.subject.keywordAuthor | Smartphone | - |
dc.subject.keywordAuthor | Smartwatch | - |
dc.subject.keywordPlus | CARDIOPULMONARY-RESUSCITATION | - |
dc.subject.keywordPlus | RESCUER FATIGUE | - |
dc.subject.keywordPlus | CPR | - |
dc.subject.keywordPlus | DEPTH | - |
dc.subject.keywordPlus | QUALITY | - |
dc.subject.keywordPlus | SMARTPHONE | - |
dc.subject.keywordPlus | GUIDELINES | - |
dc.subject.keywordPlus | MANNEQUIN | - |
dc.subject.keywordPlus | APPS | - |
dc.relation.journalResearchArea | General & Internal Medicine | - |
dc.relation.journalResearchArea | Emergency Medicine | - |
dc.relation.journalWebOfScienceCategory | Critical Care Medicine | - |
dc.relation.journalWebOfScienceCategory | Emergency Medicine | - |
dc.description.journalRegisteredClass | sci | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
84, Heukseok-ro, Dongjak-gu, Seoul, Republic of Korea (06974)02-820-6194
COPYRIGHT 2019 Chung-Ang University All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.