Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Development of an LCD-Based Visual Field System

Full metadata record
DC Field Value Language
dc.contributor.authorJoo, Jin-Ho-
dc.contributor.authorLee, Jihyoung-
dc.contributor.authorYou, Heecheon-
dc.contributor.authorKang, Jaheon-
dc.date.accessioned2024-07-10T06:00:43Z-
dc.date.available2024-07-10T06:00:43Z-
dc.date.issued2018-01-
dc.identifier.issn1011-8934-
dc.identifier.issn1598-6357-
dc.identifier.urihttps://scholarworks.bwise.kr/cau/handle/2019.sw.cau/74682-
dc.description.abstractBackground: The present study investigated the diagnostic effectiveness of an LCD-based visual field testing system (LVF) in comparison with the standard automated perimetry Humphrey Field Analyzer II-750i (HFA). Methods: A randomized controlled crossover study was conducted with 202 normal and 128 glaucomatous eyes using both LVF and HFA. The visual field testing systems were compared in terms of mean deviation (MD), pattern standard deviation (PSD), and area under the receiver operating characteristics curve (AUC) of MD and PSD differentiating the normal and glaucomatous eyes. Results: Significant correlations were found between MD measurements from LVF and those from HFA for normal eyes (r = 0.342) and glaucomatous eyes (r = 0.796); slightly higher significant correlations were identified between PSD measurements from LVF and those from HFA for normal eyes (r = 0.363) and glaucomatous eyes (r = 0.828). Furthermore, high AUCs of MD were found as 0.786 for LVF and 0.868 for HFA and AUCs of PSD as 0.913 for LVF and 0.932 for HFA. Conclusion: The comparison results of the present study support the competence of LVF compared with HFA in visual field testing for early detection of glaucoma.-
dc.language영어-
dc.language.isoENG-
dc.publisherKOREAN ACAD MEDICAL SCIENCES-
dc.titleDevelopment of an LCD-Based Visual Field System-
dc.typeArticle-
dc.identifier.doi10.3346/jkms.2018.33.e19-
dc.identifier.bibliographicCitationJOURNAL OF KOREAN MEDICAL SCIENCE, v.33, no.3-
dc.description.isOpenAccessY-
dc.identifier.wosid000426328500004-
dc.identifier.scopusid2-s2.0-85041099200-
dc.citation.number3-
dc.citation.titleJOURNAL OF KOREAN MEDICAL SCIENCE-
dc.citation.volume33-
dc.type.docTypeArticle-
dc.publisher.location대한민국-
dc.subject.keywordAuthorLCD-based Visual Field Testing System-
dc.subject.keywordAuthorStandard Automated Perimetry-
dc.subject.keywordAuthorMean Deviation-
dc.subject.keywordAuthorPattern Standard Deviation-
dc.subject.keywordAuthorArea Under the Receiver Operating Characteristics Curve-
dc.subject.keywordPlusFREQUENCY-DOUBLING TECHNOLOGY-
dc.subject.keywordPlusSTANDARD AUTOMATED PERIMETRY-
dc.subject.keywordPlusMATRIX-
dc.subject.keywordPlusDEFECTS-
dc.relation.journalResearchAreaGeneral & Internal Medicine-
dc.relation.journalWebOfScienceCategoryMedicine, General & Internal-
dc.description.journalRegisteredClasssci-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.description.journalRegisteredClasskci-
Files in This Item
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE