Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

GaN-based 4-MHz Full-Bridge Electrosurgical Generator Using Zero-Voltage Switching over Wide Load Impedance Range

Authors
Hwang, SungwoonLee, SeokminYoon, ByeongcheolJeong, ByongchangKim, JongchanKim, SungminKim, Junghyun
Issue Date
Oct-2023
Publisher
Institute of Electrical and Electronics Engineers
Keywords
Electrosurgical generator (ESG); GaN switch; high efficiency; high-frequency full-bridge inverter; Impedance; Inverters; Power generation; resonant filter; Resonant frequency; RLC circuits; Surgery; wide load impedance range; Zero voltage switching; zero-voltage switching (ZVS)
Citation
IEEE Transactions on Biomedical Circuits and Systems, v.17, no.5, pp 1125 - 1134
Pages
10
Indexed
SCIE
SCOPUS
Journal Title
IEEE Transactions on Biomedical Circuits and Systems
Volume
17
Number
5
Start Page
1125
End Page
1134
URI
https://scholarworks.bwise.kr/erica/handle/2021.sw.erica/117747
DOI
10.1109/TBCAS.2023.3310880
ISSN
1932-4545
1940-9990
Abstract
In this study, a 4-MHz full-bridge inverter system for an electrosurgical generator (ESG) is presented with zero-voltage switching (ZVS) turn-on over a wide load impedance range based on wideband gap material gallium nitride (GaN) technology. The proposed resonant circuit is used for impedance matching and limits the current when the load is short state while performing ZVS over a wide load impedance range. It also ensures high power performance at high frequencies for electric surgery. The implementation methods of the GaN switch at 4 MHz with high output performance are guided by the low inductance from the gate-driver output to the switch gate node and the effective heating management structure of the GaN-based inverter. The proposed generator achieved a maximum output power of 99 W to the load and a maximum overall efficiency of 89% with overcurrent of <3 A at short state of the load. The ESG can generate constant output power of 20, 50, 70 W through mode control of 2, 5, and 7 s. The overcurrent Moreover, the proposed generator is tested on porcine tissue samples to verify its effectiveness as an ESG. IEEE
Files in This Item
Go to Link
Appears in
Collections
COLLEGE OF ENGINEERING SCIENCES > SCHOOL OF ELECTRICAL ENGINEERING > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Sung min photo

Kim, Sung min
ERICA 공학대학 (SCHOOL OF ELECTRICAL ENGINEERING)
Read more

Altmetrics

Total Views & Downloads

BROWSE