A Temporal Interference Stimulation Driver IC with Unified Source and Sink Current Capability
- Authors
- 김종석
- Issue Date
- Oct-2025
- Publisher
- 반도체설계교육센터
- Citation
- IDEC Journal of Integrated Circuits and Systems, v.11, no.4, pp 39 - 45
- Pages
- 7
- Indexed
- KCI
- Journal Title
- IDEC Journal of Integrated Circuits and Systems
- Volume
- 11
- Number
- 4
- Start Page
- 39
- End Page
- 45
- URI
- https://scholarworks.bwise.kr/erica/handle/2021.sw.erica/126671
- ISSN
- 2384-2113
- Abstract
- This paper presents a temporal interference stimulation driver IC (TIS-IC) for deep brain stimulation. The proposed MTIS-IC (Multi-TIS) features high linearity, independent frequency and amplitude control, and a highvoltage amplifier-based architecture. It achieves a peak-to-peak output voltage error within 0.4% and a maximum output current of 2 mA. Measured results show a maximum SFDR of 59 dB and a maximum SNDR of 46 dB. A proposed calibration method improves output signal accuracy and channel uniformity. Fabricated using a 180 nm BCD process, the proposed MTIS-IC overcomes the limitations of existing systems, enabling scalable neural stimulation and expanding biomedical applications.
- Files in This Item
- There are no files associated with this item.
- Appears in
Collections - COLLEGE OF ENGINEERING SCIENCES > SCHOOL OF ELECTRICAL ENGINEERING > 1. Journal Articles

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.