Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

고진공에서 이온 카운터를 사용한 실시간 입자 모니터링 시스템의 개발 및 성능평가

Authors
안강호김용민윤진욱권용택
Issue Date
Jul-2006
Publisher
사단법인 한국입자에어로졸학회
Citation
사단법인 한국입자에어로졸학회, 2006년도 한국입자에어로졸학회 학술대회, pp 255 - 258
Pages
4
Indexed
OTHER
Journal Title
사단법인 한국입자에어로졸학회, 2006년도 한국입자에어로졸학회 학술대회
Start Page
255
End Page
258
URI
https://scholarworks.bwise.kr/erica/handle/2021.sw.erica/44748
Abstract
In this paper, the new method which is monitoring quantity of particles using by ion-counter has been developed. ISPM system is composed by Gerdien type ion-counter (house-made), DC power supply and electrometer. Ion-counter applied positive voltage could detect only positive charged particles. Therefore charged particles to Boltzmann equilibrium distribution or to some identified charge distribution can be detected by ion-counter. Ion-counter could install on the exhaust line of process equipment since pressure loss is structurally low. ISPM system has been certified by comparison with the result of SMPS (Scanning Mobility Particle Sizer) system. The relation coefficiency is above 0.98 about 20∼300nm particles with identified charge distribution under 0.1∼10.0 Torr.