- In this paper, the new method which is monitoring quantity of particles using by ion-counter has been developed. ISPM system is composed by Gerdien type ion-counter (house-made), DC power supply and electrometer. Ion-counter applied positive voltage could detect only positive charged particles. Therefore charged particles to Boltzmann equilibrium distribution or to some identified charge distribution can be detected by ion-counter. Ion-counter could install on the exhaust line of process equipment since pressure loss is structurally low. ISPM system has been certified by comparison with the result of SMPS (Scanning Mobility Particle Sizer) system. The relation coefficiency is above 0.98 about 20∼300nm
particles with identified charge distribution under 0.1∼10.0
Torr.