The effect of tube voltage combination on image artefact and radiation dose in dual-source dual-energy CT: comparison between conventional 80/140 kV and 80/150 kV plus tin filter for gout protocol

Authors
Jeon, Ji YoungLee, Sheen-WooJeong, Yu MiBaek, Han Joo
Issue Date
Mar-2019
Publisher
SPRINGER
Keywords
Gout; Tomography; X-ray computed; Tin; Artefacts; Radiation dosage
Citation
EUROPEAN RADIOLOGY, v.29, no.3, pp.1248 - 1257
Journal Title
EUROPEAN RADIOLOGY
Volume
29
Number
3
Start Page
1248
End Page
1257
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/1754
DOI
10.1007/s00330-018-5622-9
ISSN
0938-7994
Abstract
Objectives: In dual-source CT, dual-energy (DE) performance is affected by various X-ray tube voltage combinations with and without tin filter (Sn). The purpose of this study was to assess the utility of the 80/150 Sn kV voltage combination in terms of image artefact and radiation dose for DECT gout protocol, compared with the conventional 80/140 kV. Methods: Seventy-four patients with suspected gout who underwent dual-source DECT examinations scanned at 80/140 kV (n = 37) and at 80/150 Sn kV (n = 37) were included. Patients' age, sex, and serum uric acid levels were matched between the two groups. The types and incidence of image artefacts and radiation dose were evaluated. Results: The 80/150 Sn kV group had significantly fewer patients with artefacts, compared to the 80/140 kV group [11 (30 %) of 37 vs 35 (94.6 %) of 37, p < 0.001]. Except for the motion artefact, the rest of the artefactsskin, nail bed, submillimetre, motion, vascular, beam-hardening, clumpy artefact along tendonwere significantly less observed in the 80/150 Sn kV acquisitions. The dose-length product (DLP) and effective dose were significantly lower for the 80/150 Sn kV acquisitions compared with the 8s0/140 kV scans (DLP: 104.46 10.66 mGy<bold>cm vs </bold>344.70 56.39 mGy<bold>cm</bold>, p < 0.001; effective dose: 1.04 0.11 mSv vs 3.45 0.56 mSv, p < 0.001). Conclusions: The 80/150 Sn kV voltage combination in dual-source DECT system could be used as one of the artefact reduction methods while reducing radiation dose for gout protocol when compared to the conventional 80/140 kV. Key Points: DECT has emerged as the leading modality for non-invasive diagnosis of gout. Various X-ray tube voltage combinations are now feasible in dual-source DECT. The 80/150 Sn kV acquisition could facilitate artefact reduction in gout protocol.
Files in This Item
There are no files associated with this item.
Appears in
Collections
의과대학 > 의학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Jeon, Ji young photo

Jeon, Ji young
College of Medicine (Department of Medicine)
Read more

Altmetrics

Total Views & Downloads

BROWSE