Development and clinical study of the use of infrared radiation to accelerate the dissolution rate of a microneedle array patch (MAP)

Authors
Park, Jee-HyunKim, Cheong BiLee, Hyun-JiRoh, Joo YoungLee, Jae MyunKim, Hee JooPark, Jung-Hwan
Issue Date
Jun-2020
Publisher
SPRINGER HEIDELBERG
Keywords
Active evaporating; Clinical study; Dissolution rate; Dissolving microneedle array patch (MAP); Erythema; Infrared radiation; Transepidermal water loss (TEWL)
Citation
Drug Delivery and Translational Research, v.10, no.3, pp.791 - 800
Journal Title
Drug Delivery and Translational Research
Volume
10
Number
3
Start Page
791
End Page
800
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/17738
DOI
10.1007/s13346-020-00710-x
ISSN
2190-393X
Abstract
The dissolution rate of a microneedle array patch (MAP) determines how long a MAP must remain attached to the skin (often called “wear time”). In this study, the dissolution rate of a MAP was increased, not by changing the drug formulation but by employing an infrared (IR) device that is widely used for hospital treatment and in-home therapy. A MAP with microneedles 480 μm in height was prepared using hyaluronic acid (HA). Changes in transepidermal water loss (TEWL), the surface temperature of the skin, and the dissolution rate of the MAP tips with IR irradiation were evaluated on human skin in vivo. Time for recovery from erythema that occurred after MAP attachment and IR irradiation was also evaluated. TEWL increased more than fourfold with IR irradiation. Water that evaporated as a result of IR irradiation was trapped in the skin layer by the patch, resulting in the increased dissolution rate of the MAP tips. After 10 min of IR irradiation, the height of the dissolving tips compared with their initial height increased from 41 to 56%, and the dissolved volume of the tips compared with their initial volume increased from 7 to 18%. During the 10 min of irradiation, the skin surface temperature rose from 32 to 40 °C. Erythema occurred in the early stage of treatment with IR irradiation and MAP attachment, but it abated within 2 h after removal of the MAP and cessation of IR irradiation. Through this study, it was possible to shorten the administration time of MAPs by using an IR device that could be easily accessed. This method can be applied to various types of MAPs in order to reduce the time that the MAPs must remain attached to the skin without changing the drug formulation. [Figure not available: see fulltext.]. © 2020, Controlled Release Society.
Files in This Item
There are no files associated with this item.
Appears in
Collections
바이오나노대학 > 바이오나노학과 > 1. Journal Articles
의과대학 > 의학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Park, Jung Hwan photo

Park, Jung Hwan
BioNano Technology (Department of BioNano Technology)
Read more

Altmetrics

Total Views & Downloads

BROWSE