Detailed Information

Cited 5 time in webofscience Cited 6 time in scopus
Metadata Downloads

Lock-in Amplifier-Based Impedance Detection of Tissue Type Using a Monopolar Injection Needle

Authors
Kim, JunsubAbbasi, Muhammad AitzazKim, TaeheePark, Ki DeokCho, Sungbo
Issue Date
Nov-2019
Publisher
MDPI
Keywords
bio-impedance; intra-articular injection therapy; lock-in amplifier; monopolar injection needle; porcine tissue
Citation
SENSORS, v.19, no.21
Journal Title
SENSORS
Volume
19
Number
21
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/17990
DOI
10.3390/s19214614
ISSN
1424-8220
Abstract
For successful intra-articular injection therapy, it is essential to accurately position the tip of the injection needle into the target joint area while administering the drug into the affected tissue. In this study, we investigated the feasibility of a monopolar injection needle and lock-in amplifier (LIA)-based impedance measurement system for detecting the tissue type where the needle tip is located. After positioning the monopolar injection needle tip into the dermis, hypodermis, or muscle layer of pork tissue, the electrical impedance was measured in the frequency range of 10 Hz to 10 kHz. We observed a difference in the results based on the tissue type where the needle was positioned (p-value < 0.01). Therefore, the monopolar injection needle with electrical impedance measurement can be used to improve intra-articular injection therapy through non-destructive and real-time monitoring of the needle position in the tissues.
Files in This Item
There are no files associated with this item.
Appears in
Collections
IT융합대학 > 전자공학과 > 1. Journal Articles
의과대학 > 의학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Park, Ki Deok photo

Park, Ki Deok
College of Medicine (Department of Medicine)
Read more

Altmetrics

Total Views & Downloads

BROWSE