Over-the-wire deployment techniques of option elite inferior vena cava filter: 3D printing vena cava phantom study

Authors
Park B.G.Seo A.Lee S.Y.Cha J.G.Hong J.Lee H.Heo J.Do Y.W.
Issue Date
Mar-2020
Publisher
Elsevier Ltd
Keywords
3D printing vena cava phantom; Filter jumping; Filter tilt; Option IVC filter; Over-the-wire deployment system
Citation
European Journal of Radiology Open, v.7
Journal Title
European Journal of Radiology Open
Volume
7
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/26343
DOI
10.1016/j.ejro.2020.100227
ISSN
2352-0477
Abstract
Purpose: To compare filter tilt and filter jumping during Option inferior vena cava (IVC) filter deployment with 3 different wires techniques using a 3-dimensional (3D) printing vena cava phantom. Materials and methods: An IVC 3D printed vena cava phantom was made from a healthy young male's computed tomographic data. Option IVC filters were deployed with 3 different wires: i) original push wire, ii) hydrophilic stiff wire, and iii) bent stiff wire. Right internal jugular and right femoral access were used 5 times with each wire. Filter tilt angle, tilt ratio, jumping, and tip abutment to the IVC wall were analyzed. Results: The transfemoral approach with original push wire had significantly higher tilt angle than did the transjugular approach (6.1˚ ± 1.9 vs. 3.5˚ ± 1.3, p = 0.04). Mean tilt ratio was significantly lower with the bent wire with transfemoral access (0.49 ± 0.13 vs. 0.78 ± 0.18 [original push-wire] and 0.67 ± 0.08 [stiff wire], p = 0.019). The ratio was lower also with original push wire with transjugular access (0.34 ± 0.19 vs. 0.57 ±0.11 [stiff wire] and 0.58 ±0.17 [bent wire], p = 0.045). Filter jumping occurred more often with the transjugular approach with original push wire than with stiff or bent-wire delivery. Filter tip abutment to the IVC wall occurred only with the transfemoral approach. Conclusions: Bent wire with transfemoral access and original push wire with transjugular access had lower filter tilt ratio at Option IVC filter deployment. However, filter jumping was common using the original push wire with transjugular access. © 2020
Files in This Item
There are no files associated with this item.
Appears in
Collections
ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE