Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

140 keV 감마선 차폐 시 납 차폐체 두께에 따른 깊이별 선량 평가Shielding 140 keV Gamma Ray Evaluation of Dose by Depth According to Thickness of Lead Shield

Other Titles
Shielding 140 keV Gamma Ray Evaluation of Dose by Depth According to Thickness of Lead Shield
Authors
김지영이왕희안성민
Issue Date
2018
Publisher
대한방사선과학회
Keywords
99mTc; 140 keV 감마선; 표층선량; 심부선량; 방사선차폐; 99mTc; 140 keV Gamma ray; Shallow Dose; Depth Dose; Radiation Shield
Citation
방사선기술과학, v.41, no.2, pp.129 - 134
Journal Title
방사선기술과학
Volume
41
Number
2
Start Page
129
End Page
134
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/5092
DOI
10.17946/JRST.2018.41.2.129
ISSN
2288-3509
Abstract
The present study made a phantom for gamma ray of 140 keV radiated from 99mTc, examined shielding effect of lead by thickness of the shielding material, and measured surface dose and depth dose by body depth. The OSL Nano Dot dosimeter was inserted at 0, 3, 15, 40, 90, and 180㎜ depths of the phantom, and when there was no shield, 0.2㎜ lead shield, 0.5㎜ lead shield, The depth dose was measured. Experimental results show that the total cumulative dose of dosimeters with depth is highest at 366.24 uSv without shield and lowest at 94.12 uSv with 0.5㎜ lead shield. The shielding effect of 0.2㎜ lead shielding was about 30.18% and the shielding effect of 0.5㎜ lead shielding was 74.30%, when the total sum of the accumulated doses of radiation dosimeter was 100%. The phantom depth and depth dose measurements showed the highest values at 0㎜ depth for all three experiments and the dose decreases as the depth increases. This study proved that the thicker a shielding material, the highest its shielding effect is against gamma ray of 140 keV. However, it was known that shielding material can't completely shield a body from gamma ray; it reached deep part of a human body. Aside from the International Commission on Radiation Units and Measurements (ICRU) recommending depth dose by 10㎜ in thickness, a plan is necessary for employees working in department of nuclear medicine where they deal with gamma ray, which is highly penetrable, to measure depth dose by body depth, which can help them manage exposed dose properly.
Files in This Item
There are no files associated with this item.
Appears in
Collections
보건과학대학 > 방사선학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Ahn, Sung Min photo

Ahn, Sung Min
Health Science (Dept.of Radiology)
Read more

Altmetrics

Total Views & Downloads