Detailed Information

Cited 4 time in webofscience Cited 5 time in scopus
Metadata Downloads

Optimization of mask size with median modified Wiener filter algorithm for gamma images using pixelated semiconductor detector: Monte Carlo simulation study

Authors
Park, C.R.Kang, S.-H.Lee, Y.
Issue Date
Nov-2020
Publisher
Elsevier B.V.
Keywords
Gamma camera image; Median modified Wiener filter; Monte Carlo simulation; Noise reduction algorithm; Pixelated semiconductor detector
Citation
Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, v.980
Journal Title
Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume
980
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/78717
DOI
10.1016/j.nima.2020.164472
ISSN
0168-9002
Abstract
Gamma images inherently contain noise distributions owing to the low photon count in a detector. Several methods using semiconductor detectors have been reported to overcome this challenge. Median modified Wiener filter (MMWF) algorithm is one of the effective noise reduction methods, which employs mask size as an essential parameter for optimization. In this study, we evaluated and optimized the acquired gamma image using the proposed MMWF algorithm based on the Monte Carlo simulation tool with various mask sizes (3 × 3, 5 × 5, and 7 × 7) in a cadmium telluride pixelated semiconductor detector. Our results demonstrated that the images obtained with 7 × 7 mask size showed the lowest values in the normalized noise power spectrum. The contrast-to-noise ratio and the coefficient of variation were observed to be 29.2–65.9% better for the images obtained with the 7 × 7 mask size compared with those obtained using other mask sizes. This study suggests that an excellent image quality of gamma images can be achieved with semiconductor detectors by optimizing the mask size in the MMWF algorithm. © 2020
Files in This Item
There are no files associated with this item.
Appears in
Collections
보건과학대학 > 방사선학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Youngjin photo

Lee, Youngjin
Health Science (Dept.of Radiology)
Read more

Altmetrics

Total Views & Downloads

BROWSE