Estimation of Intraocular Lens Power Calculation after Myopic Corneal Refractive Surgery: Using Corneal Height in Anterior Segment Optical Coherence Tomography

Authors
김동현김미금위원량
Issue Date
Jun-2015
Publisher
대한안과학회
Keywords
Key Words: Anterior segment optical coherence tomography; Corneal height; Corneal refractive surgery; Effective lens position; Intraocular lens power calculation
Citation
Korean Journal of Ophthalmology, v.29, no.3, pp.195 - 202
Journal Title
Korean Journal of Ophthalmology
Volume
29
Number
3
Start Page
195
End Page
202
URI
https://scholarworks.bwise.kr/gachon/handle/2020.sw.gachon/82194
DOI
10.3341/kjo.2015.29.3.195
ISSN
1011-8942
Abstract
Purpose: To investigate the feasibility of estimating effective lens position (ELP) and calculating intraocular lens power using corneal height (CH), as measured using anterior segment optical coherence tomography (ASOCT), in patients who have undergone corneal refractive surgery. Methods: This study included 23 patients (30 eyes) who have undergone myopic corneal refractive surgery and subsequent successful cataract surgery. The CH was measured with AS-OCT, and the measured ELP (ELPm) was calculated. Intraocular lens power, which could achieve actual emmetropia (Preal), was determined with medical records. Estimated ELP (ELPest) was back-calculated using Preal, axial length, and keratometric value through the SRK/T formula. After searching the best-fit regression formula between ELPm and ELPest, converted ELP and intraocular lens power (ELPconv, Pconv) were obtained and then compared to ELPest and Preal, respectively. The proportion of eyes within a defined error was investigated. Results: Mean CH, ELPest, and ELPm were 3.71 ± 0.23, 7.74 ± 1.09, 5.78 ± 0.26 mm, respectively. The ELPm and ELPest were linearly correlated (ELPest = 1.841 × ELPm – 2.018, p = 0.023, R = 0.410) and ELPconv and Pconv agreed well with ELPest and Preal, respectively. Eyes within ±0.5, ±1.0, ±1.5, and ±2.0 diopters of the calculated Pconv, were 23.3%, 66.6%, 83.3%, and 100.0%, respectively. Conclusions: Intraocular lens power calculation using CH measured with AS-OCT shows comparable accuracy to several conventional methods in eyes following corneal refractive surgery.
Files in This Item
There are no files associated with this item.
Appears in
Collections
의과대학 > 의학과 > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Dong Hyun photo

Kim, Dong Hyun
College of Medicine (Department of Medicine)
Read more

Altmetrics

Total Views & Downloads

BROWSE